Package: Lecroy_Waverunner_2_Series_1_Year_IEEE_9500-R1.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Lecroy LT264/M/L Waverunner2 (1 Yr):IEEE/9500-5hd :1.00, Revision: 1
Lecroy LT354/M/L Waverunner2 (1 Yr):IEEE/9500-5hd :1.00, Revision: 1
Lecroy LT374/M/L Waverunner2 (1 Yr):IEEE/9500-5hd :1.00, Revision: 1
Lecroy LT262 Waverunner2 (1 Yr):IEEE/9500-3hd :1.00, Revision: 1
Lecroy LT372 Waverunner2 (1 Yr):IEEE/9500-3hd :1.00, Revision: 1
Lecroy LT262 Waverunner2 (1 Yr):IEEE/9500-1hd :1.00, Revision: 1
Lecroy LT264/M/L Waverunner2 (1 Yr):IEEE/9500-1hd :1.00, Revision: 1
Lecroy LT354/M/L Waverunner2 (1 Yr):IEEE/9500-1hd :1.00, Revision: 1
Lecroy LT372 Waverunner2 (1 Yr):IEEE/9500-1hd :1.00, Revision: 1
Lecroy LT374/M/L Waverunner2 (1 Yr):IEEE/9500-1hd :1.00, Revision: 1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Lecroy LT264/M/L Waverunner2 (1 Yr):IEEE/9500-5hd :1.00
Lecroy LT354/M/L Waverunner2 (1 Yr):IEEE/9500-5hd :1.00
Lecroy LT374/M/L Waverunner2 (1 Yr):IEEE/9500-5hd :1.00
Date: 2001-09-06
Revision: 1
Automated Reference: Datron 9500
Automated Reference: Datron 9510
Lecroy LT262 Waverunner2 (1 Yr):IEEE/9500-3hd :1.00
Lecroy LT372 Waverunner2 (1 Yr):IEEE/9500-3hd :1.00
Date: 2001-09-06
Revision: 1
Automated Reference: Datron 9500
Automated Reference: Datron 9510
Procedure ID No: 001240M002
Models covered: LT262 LT372
Source: Lecroy Digital Waverunner 2
Service Manual
January 2001
Rev A
Source Author: N/K
Compatibility: MET/CAL 6.10 or later
Subprocedures:
Sub Waverunner 2 (1 yr) Leakage Current IEEE : 1.00
Sub Waverunner 2 (1 yr) Input Noise IEEE : 1.00
Sub Datron 9500-5 Channel Selection: 1.00
Sub Waverunner 2 (1 yr) Impedance IEEE/9500 : 1.00
Sub UUT Four Channel Select/: 1.00
Sub Waverunner 2 (1 yr) DC Accuracy IEEE/9500 :1.00
Sub Waverunner 2 (1 yr) Bandwidth IEEE /9500 : 1.00
Sub Lecroy LT Series (1 yr) Risetime IEEE /9500 : 1.00
Sub Waverunner 2 (1 yr) Trig Level IEEE/9500 : 1.00
Sub Waverunner 2 (1 yr) Smart Trig IEEE/9500 : 1.00
Sub Waverunner 2 (1 yr) Timebase IEEE /9500 : 1.00
Required Files:
None
Additional equipment Required:
One Digital Multimeter : Accuracy 0.1 %
The Procedure was tested using:
LECROY Waverunner2 LT374L
Serial Number: LT37400158
Soft Version : LT374L 08.9.0
Soft Options : DFP JTA PMT WAVA
Hard Options : CPUP VGAC GPIB R232 CLBZ
FD01 CENT NET I2C GP02
The average run time during testing was 25 minutes.
Procedure history:
22-08-01: First release as 9500 procedure by MSD,
Fluke Norwich
The procedure is intended for use with Met/Cal(r) calibration software,
and the terms and conditions set forth in your METCAL license apply to
this procedure.
Due to our policy of continuously up dating our products, this procedure
may contain minor differences in specification to those found in the manual
or other documentation. While every effort has been made to ensure that
this procedure is accurate, Fluke cannot be held responsible for the
consequences of error or omissions found within this procedure.
The copyright in this procedure is owned by Fluke Precision Measurement
Limited.
NUMERIC REGISTERS USED:
M[100]
This register specifies which channel of the oscilloscope
is currently being tested.
M[100] = 1 means Channel 1
M[100] = 2 means Channel 2
M[100] = 3 means Channel 3
M[100] = 4 means Channel 4
M[103]
This register specifies which LT Series Model
is currently is being tested.
M[102]
This register specifies the Bandwidth of LT Series Model
being tested.
M[101]
This register specifies the number of channels
LT Series Model has.
Lecroy LT262 Waverunner2 (1 Yr):IEEE/9500-1hd :1.00
Lecroy LT264/M/L Waverunner2 (1 Yr):IEEE/9500-1hd :1.00
Lecroy LT354/M/L Waverunner2 (1 Yr):IEEE/9500-1hd :1.00
Lecroy LT372 Waverunner2 (1 Yr):IEEE/9500-1hd :1.00
Lecroy LT374/M/L Waverunner2 (1 Yr):IEEE/9500-1hd :1.00
Date: 2001-09-06
Revision: 1
Automated Reference: Datron 9500
Automated Reference: Datron 9510
Procedure ID No: 001240M003
Models covered: LT262 LT264 LT354 LT372 LT374
Source: Lecroy Digital Waverunner 2
Service Manual
January 2001
Rev A
Source Author: N/K
Compatibility: MET/CAL 6.10 or later
Subprocedures:
Sub Waverunner 2 (1 yr) Leakage Current IEEE : 1.00
Sub Waverunner 2 (1 yr) Input Noise IEEE : 1.00
Sub Datron 9500-5 Channel Selection: 1.00
Sub Waverunner 2 (1 yr) Impedance IEEE/9500 : 1.00
Sub UUT Four Channel Select/: 1.00
Sub Waverunner 2 (1 yr) DC Accuracy IEEE/9500 :1.00
Sub Waverunner 2 (1 yr) Bandwidth IEEE /9500 : 1.00
Sub Lecroy LT Series (1 yr) Risetime IEEE /9500 : 1.00
Sub Waverunner 2 (1 yr) Trig Level IEEE/9500-1hd : 1.00
Sub Waverunner 2 (1 yr) Smart Trig IEEE/9500 : 1.00
Sub Waverunner 2 (1 yr) Timebase IEEE /9500 : 1.00
Required Files:
None
Additional equipment Required:
One Digital Multimeter : Accuracy 0.1 %
The Procedure was tested using:
LECROY Waverunner2 LT374L
Serial Number: LT37400158
Soft Version : LT374L 08.9.0
Soft Options : DFP JTA PMT WAVA
Hard Options : CPUP VGAC GPIB R232 CLBZ
FD01 CENT NET I2C GP02
The average run time during testing was 45 minutes.
Procedure history:
22-08-01: First release as 9500 procedure by MSD,
Fluke Norwich
The procedure is intended for use with Met/Cal(r) calibration software,
and the terms and conditions set forth in your METCAL license apply to
this procedure.
Due to our policy of continuously up dating our products, this procedure
may contain minor differences in specification to those found in the manual
or other documentation. While every effort has been made to ensure that
this procedure is accurate, Fluke cannot be held responsible for the
consequences of error or omissions found within this procedure.
The copyright in this procedure is owned by Fluke Precision Measurement
Limited.
NUMERIC REGISTERS USED:
M[100]
This register specifies which channel of the oscilloscope
is currently being tested.
M[100] = 1 means Channel 1
M[100] = 2 means Channel 2
M[100] = 3 means Channel 3
M[100] = 4 means Channel 4
M[103]
This register specifies which LT Series Model
is currently is being tested.
M[102]
This register specifies the Bandwidth of LT Series Model
being tested.
M[101]
This register specifies the number of channels
LT Series Model has.
Procedure ID No: 001240M001
Models covered: LT264 LT354 LT374
Source: Lecroy Digital Waverunner 2
Service Manual
January 2001
Rev A
Source Author: N/K
Compatibility: MET/CAL 6.10 or later
Subprocedures:
Sub Waverunner 2 (1 yr) Leakage Current IEEE : 1.00
Sub Waverunner 2 (1 yr) Input Noise IEEE : 1.00
Sub Datron 9500-5 Channel Selection: 1.00
Sub Waverunner 2 (1 yr) Impedance IEEE/9500 : 1.00
Sub UUT Four Channel Select/: 1.00
Sub Waverunner 2 (1 yr) DC Accuracy IEEE/9500 :1.00
Sub Waverunner 2 (1 yr) Bandwidth IEEE /9500 : 1.00
Sub Lecroy LT Series (1 yr) Risetime IEEE /9500 : 1.00
Sub Waverunner 2 (1 yr) Trig Level IEEE/9500 : 1.00
Sub Waverunner 2 (1 yr) Smart Trig IEEE/9500 : 1.00
Sub Waverunner 2 (1 yr) Timebase IEEE /9500 : 1.00
Required Files:
None
Additional equipment Required:
One Digital Multimeter : Accuracy 0.1 %
The Procedure was tested using:
LECROY Waverunner2 LT374L
Serial Number: LT37400158
Soft Version : LT374L 08.9.0
Soft Options : DFP JTA PMT WAVA
Hard Options : CPUP VGAC GPIB R232 CLBZ
FD01 CENT NET I2C GP02
The average run time during testing was 50 minutes.
Procedure history:
22-08-01: First release as 9500 procedure by MSD,
Fluke Norwich
The procedure is intended for use with Met/Cal(r) calibration software,
and the terms and conditions set forth in your METCAL license apply to
this procedure.
Due to our policy of continuously up dating our products, this procedure
may contain minor differences in specification to those found in the manual
or other documentation. While every effort has been made to ensure that
this procedure is accurate, Fluke cannot be held responsible for the
consequences of error or omissions found within this procedure.
The copyright in this procedure is owned by Fluke Precision Measurement
Limited.
NUMERIC REGISTERS USED:
M[100]
This register specifies which channel of the oscilloscope
is currently being tested.
M[100] = 1 means Channel 1
M[100] = 2 means Channel 2
M[100] = 3 means Channel 3
M[100] = 4 means Channel 4
M[103]
This register specifies which LT Series Model
is currently is being tested.
M[102]
This register specifies the Bandwidth of LT Series Model
being tested.
M[101]
This register specifies the number of channels
LT Series Model has.