Package: aemc_500_1yr_ver_9100_opt200.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
AEMC 500: (1 year) CAL VER /9100+Opt200, Revision: 1.2
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
AEMC 500: (1 year) CAL VER /9100+Opt200
Date: 2009-12-30 14:33:56
Revision: 1.2
Automated Reference: Datron 9100
Automated Reference: Datron 9100 Opt200
Automated Reference: Fluke 5500A
Automated Reference: Fluke 5500A/COIL
Model(s) Covered:
500
This procedure contains the following test(s) whose Test Uncertainty Ratio
are less than 4:1.
AC CURRENT
36 A, 50 Hz; TUR 3.2407
200 A, 440 Hz; TUR 3.0653
360 A, 50 Hz; TUR 3.9979
Source:
AEMC Model 500 Clamp-on Meter General Specifications
Compatibility:
MET/CAL V7.20 SP1 (7.20x) or later
Additional Equipment Required:
None
The average run time during testing was 6 minutes or less.