Agilent 34405A: ( 1 year) CAL VER VISA /5700,5725,FGEN

Date: 2013-02-27 08:25:05

Revision: 1.3

Automated Reference: Fluke 5700A

Automated Reference: Fluke 5725A

Automated Reference: Function Generator

Model Covered:

34405A

Source:

Agilent 34405A 5 1/2 Digit Multimeter User's and Service Guide

Part Number: 34405-91000, Fourth Edition. March 23, 2009

Compatibility:

MET/CAL 7.20 SP3 (7.20Z) or later

Flexible Standards Required:

Function Generator (FGEN):

Output Range: 1 V, 10 kHz Sine

Accuracy: 0.6 Hz or better

Example: Fluke 395 or equivalent

Additional Equipment Required:

Shorting Plug--Dual banana plug with copper wire short

between the two terminals

The average run time during testing was:

12 minutes (Root Difference Square guardbanding method used (TAR))

35 minutes (Measurement Uncertainty guardbanding method used (MU))