Package: agilent_34405a_1yr_ver_visa_5700_5725_fgen.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Agilent 34405A: ( 1 year) CAL VER VISA /5700,5725,FGEN, Revision: 1.2
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Agilent 34405A: ( 1 year) CAL VER VISA /5700,5725,FGEN
Date: 2011-09-02 13:40:10
Revision: 1.2
Automated Reference: Fluke 5700A
Automated Reference: Fluke 5725A
Automated Reference: Function Generator
Model Covered:
34405A
Source:
Agilent 34405A 5 1/2 Digit Multimeter User's and Service Guide
Part Number: 34405-91000, Fourth Edition. March 23, 2009
Compatibility:
MET/CAL 7.20 SP3 (7.20Z) or later
Flexible Standards Required:
Function Generator (FGEN):
Output Range: 1 V, 10 kHz Sine
Accuracy: 0.6 Hz or better
Example: Fluke 395 or equivalent
Additional Equipment Required:
Shorting Plug--Dual banana plug with copper wire short
between the two terminals
The average run time during testing was:
12 minutes (Root Difference Square guardbanding method used (TAR))
35 minutes (Measurement Uncertainty guardbanding method used (MU))