Agilent 34411A: (1 year) CAL VER IEEE /5720,FGEN

Date: 2012-07-03 14:17:35

Revision: 1.1

Automated Reference: Fluke 5720A

Automated Reference: Function Generator

Model Covered:

34411A

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1.

AC VOLTAGE

- 1 V, 10 Hz; TUR 2.9062

- 10 V, 10 Hz; TUR 2.8219

AC CURRENT

- 100 µA, 5 kHz; TUR 1.2727

- 1 mA, 10 kHz; TUR 1.2632

- 10 mA, 10 kHz; TUR 1.4118

- 100 mA, 10 kHz; TUR 1.9672

- 1 A, 5 kHz; TUR 2.5000

Source:

Agilent 34410A/11A 6½ Digit Multimeter (includes the L4411A 1U DMM)

Service Guide, Part No. 34410-90010, Fourth Edition, February 2007

Compatibility:

MET/CAL 7.3 SP1 (7.3.36) or later

Additional Equipment Required:

Precision 4-wire short with standard ¾" banana spacing, 2 ea.,

Agilent Model 34172A DMM Calibration Short or equivalent

Cable, 50 Ohm, BNC(m)-(m), Pomona 2249-C-36

BNC female to dual banana plug, Pomona 1269

The average run time during testing was 23 minutes (TAR).