Agilent 34972A: (1 year) CAL VER VISA /5520

Date: 2013-02-01 09:29:28

Revision: 1.3

Automated Reference: Fluke 5520A

Models covered:

Agilent 34972A w/34901A 20 Channel Multiplexer module

The following test points have a Test Uncertainty Ratio (TUR) of less than

4 to 1 based 5520A 1 year specifications.

DC VOLTS, 100 mV Range : ±100 mV : TUR 3.0000

DC VOLTS, 1 V Range : ±1 V : TUR 3.6154

DC VOLTS, 10 V Range : ±10 V : TUR 2.8571

DC VOLTS, 100 V Range : ±100 V : TUR 2.6154

AC VOLTS, 100 mV Range : 100 mV @ 50 kHz : TUR 3.9535

AC VOLTS, 1 V Range : 10 V @ 20 Hz : TUR 2.8571

AC VOLTS, 10 V Range : 10 V @ 10 Hz : TUR 2.7397

OHMS 2 WIRE, 100 Ohm Range : 100 Ohms : TUR 3.3333

OHMS 2 WIRE, 1 kOhm Range : 1 kOhm : TUR 3.6667

OHMS 2 WIRE, 10 kOhm Range : 10 kOhm : TUR 3.6667

OHMS 2 WIRE, 100 kOhm Range : 100 kOhm : TUR 3.6667

OHMS 2 WIRE, 1 MOhm Range : 1 MOhm : TUR 3.2353

OHMS 2 WIRE, 10 MOhm Range : 10 MOhms : TUR 3.0370

OHMS 4 WIRE, 100 Ohm Range : 100 Ohms : TUR 3.3333

OHMS 4 WIRE, 1 kOhm Range : 1 kOhm : TUR 3.6667

OHMS 4 WIRE, 10 kOhm Range : 10 kOhm : TUR 3.6667

OHMS 4 WIRE, 100 kOhm Range : 100 kOhm : TUR 3.6667

OHMS 4 WIRE, 1 MOhm Range : 1 MOhm : TUR 3.2353

OHMS 4 WIRE, 10 MOhm Range : 10 MOhms : TUR 3.0370

AC CURRENT, 10 mA Range : 10 mA @ 1 kHz : TUR 2.3333

AC CURRENT, 1 A Range : 1 A @ 1 kHz : TUR 2.3333

Compatibility:

MET/CAL 7.3 SP1 (7.3.36) or later

Required Equipment:

HP/Agilent 34901A 20 Channel Multiplexer module

Copper wire, insulated, ~15 cm long

Leads, Teflon®, twisted pair, shielded, 3 ea. sufficiently long to

connect UUT to source

The average run time during testing was:

15 minutes (Root Difference Square guardbanding method Mode (TAR))

28 minutes (Measurement Uncertainty guardbanding method Mode (MU))