Agilent 34980A: (1 year) CAL VER IEEE /5520, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Agilent 34980A: (1 year) CAL VER IEEE /5520
Date: 2010-11-11 06:10:41
Revision: 1.1
Automated Reference: Fluke 5520A
Models covered:
34980A
The following test points have a Test Uncertainty Ratio (TUR) of less than
4 to 1 based 5520A 1 year specifications.
DC VOLTAGE, 100 mV Range : 100 mV : TUR 3.0000
DC VOLTAGE, 100 mV Range : -100 mV : TUR 3.0000
DC VOLTAGE, 1 V Range : 1 V : TUR 3.6154
DC VOLTAGE, 1 V Range : -1 V : TUR 3.6154
DC VOLTAGE, 10 V Range : 10 V : TUR 2.8571
DC VOLTAGE, 10 V Range : -10 V : TUR 2.8571
DC VOLTAGE, 100 V Range : 100 V : TUR 2.6154
DC VOLTAGE, 100 V Range : -100 V : TUR 2.6154
AC VOLTAGE, 100 mV Range : 100 mV @ 50 kHz : TUR 3.9535
AC VOLTAGE, 10 V Range : 1 V @ 20 Hz : TUR 2.8571
AC VOLTAGE, 10 V Range : 10 V @ 10 Hz : TUR 2.7397
OHMS 2 WIRE, 100 Ohm Range : 100 Ohm : TUR 3.3333
OHMS 2 WIRE, 100 Ohm Range : 1 kOhm : TUR 3.6667
OHMS 2 WIRE, 100 Ohm Range : 10 kOhm : TUR 3.6667
OHMS 2 WIRE, 100 Ohm Range : 1 MOhm : TUR 3.3529
OHMS 2 WIRE, 100 Ohm Range : 10 MOhm : TUR 3.0400
OHMS 4 WIRE, 100 Ohm Range : 100 Ohm : TUR 3.3333
OHMS 4 WIRE, 100 Ohm Range : 1 kOhm : TUR 3.6667
OHMS 4 WIRE, 100 Ohm Range : 10 kOhm : TUR 3.6667
OHMS 4 WIRE, 100 Ohm Range : 100 kOhm : TUR 3.6667
OHMS 4 WIRE, 100 Ohm Range : 1 MOhm : TUR 3.2353
OHMS 4 WIRE, 100 Ohm Range : 10 MOhm : TUR 3.0370
AC CURRENT, 10 mA Range : 10 mA @ 1 kHz : TUR 2.3333
AC CURRENT, 1 A Range : 1 A @ 1 kHz : TUR 2.3333
Required Equipment:
See "Sub Agilent 34980A (1 yr) VER Required Equipment"
The average run time during testing was:
10 minutes (Root Difference Square guardbanding method Mode (TAR))
15 minutes (Measurement Uncertainty guardbanding method Mode (MU))