Agilent 34980A: (1 year) CAL VER IEEE /5720,FGEN

Date: 2012-10-04 07:38:45

Revision: 1.2

Automated Reference: Fluke 5720A

Automated Reference: Function Generator

Models covered:

34980A

The following test points have a Test Uncertainty Ratio (TUR) of less than

4 to 1 based 5720A 1 year specifications.

AC VOLTAGE, 10 V Range : 10 V @ 10 Hz : TUR 2.8571

AC CURRENT, 1 A Range : 1 A @ 1 kHz : TUR 3.8889

Required Equipment:

50 Ohm BNC(m)-(m) cable

Adapter, BNC(f) to Binding Posts, Pomona 1452

50 Ohm feed thru termination

User configured test fixture

The average run time during testing was:

15 minutes (Root Difference Square guardbanding method Mode (TAR))

20 minutes (Measurement Uncertainty guardbanding method Mode (MU))