Package: agilent_dso80x04b_1yr_ver_9500_g3_9560.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Agilent DSO80204B: (1 yr) CAL VER IEEE /9500+G3,5Hd

Agilent DSO80304B: (1 yr) CAL VER IEEE /9500+G3,5Hd

Agilent DSO80404B: (1 yr) CAL VER IEEE /9500+G3,5Hd

Agilent DSO80604B: (1 yr) CAL VER IEEE /9500+G3,5Hd

Date: 2006-08-03 09:48:43

Revision: 1.1

Automated Reference: Datron 9500

Automated Reference: Datron 9510

Automated Reference: Fluke 9560

Model(s) Covered:

Agilent DSO80204B, DSO80304B, DSO80404B, DSO80604B

Compatibility:

MET/CAL 7.11 SP1 (7.11j) or later

Source(s):

Agilent 80000 Series Infiniium Oscilloscopes Service Guide,

Part No. D8104-97000, February 2006

Additional Equipment Required:

Adapter, 3.5 mm (f) SMA to Precision BNC, Agilent Part No.

54855-67604, No Substitute

Average run time during testing:

20 min. (TAR mode), 1 hr. 15 min. (MU mode)

Agilent DSO80204B: (1 yr) CAL VER IEEE /9500+G3,2Hd

Agilent DSO80304B: (1 yr) CAL VER IEEE /9500+G3,2Hd

Agilent DSO80404B: (1 yr) CAL VER IEEE /9500+G3,2Hd

Agilent DSO80604B: (1 yr) CAL VER IEEE /9500+G3,2Hd

Date: 2006-08-03 09:48:46

Revision: 1.1

Automated Reference: Datron 9500

Automated Reference: Datron 9510

Automated Reference: Fluke 9560

Model(s) Covered:

Agilent DSO80204B, DSO80304B, DSO80404B, DSO80604B

Compatibility:

MET/CAL 7.11 SP1 (7.11j) or later

Source(s):

Agilent 80000 Series Infiniium Oscilloscopes Service Guide,

Part No. D8104-97000, February 2006

Additional Equipment Required:

Adapter, 3.5 mm (f) SMA to Precision BNC, Agilent Part No.

54855-67604, No Substitute

Average run time during testing:

20 min. (TAR mode), 1 hr. 15 min. (MU mode)