Agilent DSO90254A: (1 yr) VER VISA /9500+4Hd,LO,E441x, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Agilent DSO90254A: (1 yr) VER VISA /9500+4Hd,LO,E441x
Date: 2011-02-22 09:12:53
Revision: 1.1
Automated Reference: Agilent E4412A
Automated Reference: Agilent E4418B
Automated Reference: Datron 9500
Automated Reference: Datron 9510
Automated Reference: Local Oscillator
Manual Reference: Agilent 11667A
Model(s) Covered:
Agilent DSO90254A
Compatibility:
MET/CAL 7.3 or later
Source(s):
Agilent 90000 Series Infiniium Oscilloscopes Service Guide,
Part No. 54913-97007, December 2008
Flexible Standard(s) Required:
- Local Oscillator (LO):
Example: Agilent E8257D with Option 520 or equivalent
- Frequency Range: 10 MHz to 2.5 GHz, CW
Accuracy: < ± 0.01 ppm
Settability: 10 Hz (Internal)
- Amplitude Range: -24 dBm to +16 dBm
Additional Equipment Required:
- Adapter, 3.5 mm (f) SMA to Precision BNC,
Agilent Part No. 54855-67604 (No Substitute)
- Adapter, 3.5 mm (m) to Type N (m), 1 ea.
- Coaxial cable, 3.5 mm (f) to Type N (m), DC to >= 18 GHz, 1 ea.
- Power Splitter, 2 resistor, DC to >= 18 GHz,
Agilent 11667A or equivalent
Average run time during testing:
33 min.