Agilent DSO90404A: (1 yr) VER VISA /9500/3200+5Hd,LO

Date: 2011-02-22 09:14:05

Revision: 1.1

Automated Reference: Datron 9500

Automated Reference: Datron 9510

Automated Reference: Fluke 9560

Automated Reference: Local Oscillator

Model(s) Covered:

Agilent DSO90404A

Compatibility:

MET/CAL 7.3 or later

Source(s):

Agilent 90000 Series Infiniium Oscilloscopes Service Guide,

Part No. 54913-97007, December 2008

Flexible Standard(s) Required:

- Local Oscillator (LO):

Example: Agilent E8257D Opt 520 or equivalent

- Frequency Range: 10.00002 MHz, CW

Accuracy: < ± 0.01 ppm

Settability: 10 Hz (Internal)

- Amplitude Range: 0 dBm (into 50 Ohms)

Additional Equipment Required:

Adapter, 3.5 mm (f) SMA to Precision BNC, Agilent Part No.

54855-67604, No Substitute

Average run time during testing:

33 min.