Agilent DSO90804A: (1 yr) VER VISA /9500+4Hd,LO,E441x

Date: 2011-02-22 09:13:02

Revision: 1.1

Automated Reference: Agilent E4412A

Automated Reference: Agilent E4418B

Automated Reference: Datron 9500

Automated Reference: Datron 9510

Automated Reference: Local Oscillator

Manual Reference: Agilent 11667A

Model(s) Covered:

Agilent DSO90804A

Compatibility:

MET/CAL 7.3 or later

Source(s):

Agilent 90000 Series Infiniium Oscilloscopes Service Guide,

Part No. 54913-97007, December 2008

Flexible Standard(s) Required:

- Local Oscillator (LO):

Example: Agilent E8257D with Option 520 or equivalent

- Frequency Range: 10 MHz to 8 GHz, CW

Accuracy: < ± 0.01 ppm

Settability: 10 Hz (Internal)

- Amplitude Range: -24 dBm to +16 dBm

Additional Equipment Required:

- Adapter, 3.5 mm (f) SMA to Precision BNC,

Agilent Part No. 54855-67604 (No Substitute)

- Adapter, 3.5 mm (m) to Type N (m), 1 ea.

- Coaxial cable, 3.5 mm (f) to Type N (m), DC to >= 18 GHz, 1 ea.

- Power Splitter, 2 resistor, DC to >= 18 GHz,

Agilent 11667A or equivalent

Average run time during testing:

33 min.