AT E4402B: (1 yr) CAL VER /9640-LPN,4418,HFSG,LFCTR

HP E4402B: (1 yr) CAL VER /9640-LPN,4418,HFSG,LFCTR

Date: 2012-12-06 17:01:55

Revision: 1.3

Automated Reference: Agilent E4418B

Automated Reference: Agilent E4419B

Automated Reference: Fluke 9640A

Automated Reference: Fluke 9640A-50

Automated Reference: High Frequency Sig Gen

Automated Reference: HP 8482A

Automated Reference: Low Frequency Counter

Manual Reference: Agilent 86205A

Model(s) Covered:

E4402B

Source(s):

Calibration Guide, Agilent Technologies ESA Spectrum Analyzers,

Manufacturing Part No. E4401-90493, March 2005

Compatibility:

MET/CAL 7.20 SP3 (7.20Z) or later

Flexible Standard(s) Required:

High Frequency Signal Generator (HFSG):

Frequency Range: 10 MHz to 3 GHz, CW

Frequency Accuracy: ±0.02% (Internal); External Reference capable

Amplitude Range: -35 dBm to +16 dBm

Amplitude Accuracy: ±2 dB

SSB Phase Noise: <-120 dBc/Hz @ 20 kHz offset

Example: Anritsu MG3690() or equivalent

Low Frequency Counter (LFCTR):

Ch. A Frequency Range: 10 MHz

Ch. A Frequency Accuracy: Based on external frequency standard

Example: Fluke PM 6690/6xx or equivalent

Frequency Standard:

Frequency Range: Suitable for use with configured LF counter

Frequency Accuracy: ±0.025 ppm or better (Option 1D5)

±0.625 ppm or better (standard)

Additional Equipment Required:

Splitter, Power, Range: 50 MHz to 3 GHz, Type N, 50 Ohms,

Insertion Loss: 6 dB, Port Tracking: <0.25 dB, Equivalent Output

SWR: <1.22; Agilent® 11667A or equivalent

Bridge, Directional, 5 MHz to 3 GHz, 50 Ohms, Directivity: >40 dB,

Coupling Factor: 16 dB, Insertion Loss: 2 dB max; Agilent 86205A

or equivalent

Filter, Low Pass, 300 MHz, 50 Ohms, Rejection at >435 MHz: >45 dB;

Agilent 0955-0455 or equivalent

Filter, Low Pass, 1 GHz, 50 Ohms, Rejection at >2 GHz: >60 dB;

Agilent 0955-0487 or equivalent

The average run times during testing were:

2 hr., 31 min. (TAR mode)