AT E4402B: (1 yr) CAL VER /9640,4418,HFSG,LFCTR

HP E4402B: (1 yr) CAL VER /9640,4418,HFSG,LFCTR

Date: 2008-04-01

Revision: 1.1

Automated Reference: Agilent E4418B

Automated Reference: Fluke 9640A

Automated Reference: Fluke 9640A-50

Automated Reference: HF Signal Generator

Automated Reference: HP 8482A

Automated Reference: LF Counter

Manual Reference: Agilent 86205A

Model(s) Covered:

E4402B

Source(s):

Calibration Guide, Agilent Technologies ESA Spectrum Analyzers,

Manufacturing Part No. E4401-90475, March 2005

USAF T.O. 33K3-4-3334-1, Technical Manual, Calibration Procedure

For Spectrum Analyzer, E4402B, dated 30 March 2007

Compatibility:

MET/CAL 7.20x or later

Flexible Standards Required:

HF Signal Generator (HFSG):

Frequency Range: 50 MHz to 2.56 GHz

Accuracy: ±1 kHz

Amplitude Range: -25 dBm to +5 dBm

Accuracy: NA

Phase Noise:

<-131 dBc/Hz typical @ 100 kHz offset

<-145 dBc/Hz typical @ 1 MHz offset

<-147 dBc/Hz typical @ 5 MHz offset

<-149 dBc/Hz typical @ 10 MHz offset

Example: HP 8663A or equivalent

LF Counter (LFCTR):

Frequency Range: 10 MHz

Accuracy: External frequency standard used

Accuracy: ±0.025/0.6 ppm, depending on UUT option configuration

(Internal/External frequency reference as required)

Example: Fluke PM 6690/6xx or equivalent

Additional Equipment Required:

Frequency Standard, suitable for use with configured LF counter,

±0.025 ppm or better

Termination, Type N (m), 50 Ohms, DC-3 GHz, HP/Agilent 909C or

equivalent

Power Splitter, DC-3 GHz, HP/Agilent 11667A or equivalent

Directional Bridge, 500 kHz to 3 GHz, HP/Agilent 86205A or equivalent

300 MHz Low Pass Filter, HP/Agilent 0955-0455 or equivalent

1 GHz Low Pass Filter, HP/Agilent 0955-0487 or equivalent

Adapter, precision, Type N (m)-(m), 50 Ohms, Agilent 1250-1475 or

equivalent

Cable, coaxial, 36", 50 Ohms, Type N (m) to APC 3.5 (m), Agilent

1250-1744 or equivalent

The average run times during testing were:

1 hr., 20 min. (Option 1DR equipped)