AT E4402B: (1 yr) CAL VER /9640,4418,HFSG,LFCTR, Revision: 1.1
HP E4402B: (1 yr) CAL VER /9640,4418,HFSG,LFCTR, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
AT E4402B: (1 yr) CAL VER /9640,4418,HFSG,LFCTR
HP E4402B: (1 yr) CAL VER /9640,4418,HFSG,LFCTR
Date: 2008-04-01
Revision: 1.1
Automated Reference: Agilent E4418B
Automated Reference: Fluke 9640A
Automated Reference: Fluke 9640A-50
Automated Reference: HF Signal Generator
Automated Reference: HP 8482A
Automated Reference: LF Counter
Manual Reference: Agilent 86205A
Model(s) Covered:
E4402B
Source(s):
Calibration Guide, Agilent Technologies ESA Spectrum Analyzers,
Manufacturing Part No. E4401-90475, March 2005
USAF T.O. 33K3-4-3334-1, Technical Manual, Calibration Procedure
For Spectrum Analyzer, E4402B, dated 30 March 2007
Compatibility:
MET/CAL 7.20x or later
Flexible Standards Required:
HF Signal Generator (HFSG):
Frequency Range: 50 MHz to 2.56 GHz
Accuracy: ±1 kHz
Amplitude Range: -25 dBm to +5 dBm
Accuracy: NA
Phase Noise:
<-131 dBc/Hz typical @ 100 kHz offset
<-145 dBc/Hz typical @ 1 MHz offset
<-147 dBc/Hz typical @ 5 MHz offset
<-149 dBc/Hz typical @ 10 MHz offset
Example: HP 8663A or equivalent
LF Counter (LFCTR):
Frequency Range: 10 MHz
Accuracy: External frequency standard used
Accuracy: ±0.025/0.6 ppm, depending on UUT option configuration
(Internal/External frequency reference as required)
Example: Fluke PM 6690/6xx or equivalent
Additional Equipment Required:
Frequency Standard, suitable for use with configured LF counter,
±0.025 ppm or better
Termination, Type N (m), 50 Ohms, DC-3 GHz, HP/Agilent 909C or
equivalent
Power Splitter, DC-3 GHz, HP/Agilent 11667A or equivalent
Directional Bridge, 500 kHz to 3 GHz, HP/Agilent 86205A or equivalent
300 MHz Low Pass Filter, HP/Agilent 0955-0455 or equivalent
1 GHz Low Pass Filter, HP/Agilent 0955-0487 or equivalent
Adapter, precision, Type N (m)-(m), 50 Ohms, Agilent 1250-1475 or
equivalent
Cable, coaxial, 36", 50 Ohms, Type N (m) to APC 3.5 (m), Agilent
1250-1744 or equivalent
The average run times during testing were:
1 hr., 20 min. (Option 1DR equipped)