Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
AT E4405B: (1 yr) CAL VER /9640-LPN,4418,LFCTR,LO,SWPG, Revision: 1.2
HP E4405B: (1 yr) CAL VER /9640-LPN,4418,LFCTR,LO,SWPG, Revision: 1.2
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
AT E4405B: (1 yr) CAL VER /9640-LPN,4418,LFCTR,LO,SWPG
HP E4405B: (1 yr) CAL VER /9640-LPN,4418,LFCTR,LO,SWPG
Date: 2011-10-04
Revision: 1.2
Automated Reference: Agilent E4418B
Automated Reference: Fluke 9640A
Automated Reference: Fluke 9640A-50
Automated Reference: HP 8481A
Automated Reference: LF Counter
Automated Reference: Local Oscillator
Automated Reference: Low-Frequency Counter
Automated Reference: Sweep Generator
Manual Reference: Agilent 11667A
Manual Reference: Agilent 86205A
Manual Reference: Agilent 87300B
Model(s) Covered:
E4405B
Source(s):
Calibration Guide, Agilent Technologies ESA Spectrum Analyzers,
Manufacturing Part No. E4401-90493, March 2005
Compatibility:
MET/CAL 7.20 SP3 (7.20Z) or later
Flexible Standard(s) Required:
Synthesized Sweep Generator (SWPG):
Frequency Range: 10 MHz to 14 GHz, CW
Accuracy: ±0.02% (Internal); External Reference capable
Amplitude Range: -40 dBm to +16 dBm
Resolution: 0.05 dB
SSB Phase Noise: <-80 dBc/Hz @ 50 kHz offset
Example: HP 83630B or equivalent
Local Oscillator (LO):
Frequency Range: 10 MHz to 14 GHz, CW
Accuracy: ±0.02% (Internal); External Reference capable
Amplitude Range: -40 dBm to +16 dBm
Accuracy: ±2 dB
SSB Phase Noise: <-80 dBc/Hz @ 50 kHz offset
Example: Anritsu MG3694() or equivalent
LF Counter (LFCTR):
Ch. A Frequency Range: 10 MHz
Accuracy: Based on external frequency standard
Example: Fluke PM 6690/6xx or equivalent
Frequency Standard:
Frequency Range: Suitable for use with configured LF counter
Accuracy: ±0.025 ppm or better (Option 1D5)
±0.625 ppm or better (standard)
Additional Equipment Required:
Splitter, Power, Range: 50 MHz to 13.2 GHz, Type N, 50 Ohms,
Insertion Loss: 6 dB, Port Tracking: <0.25 dB, Equivalent Output
SWR: <1.22; Agilent® 11667A or equivalent
Bridge, Directional, 5 MHz to 3 GHz, 50 Ohms, Directivity: >40 dB,
Coupling Factor: 16 dB, Insertion Loss: 2 dB max; Agilent 86205A
or equivalent
Coupler, Directional, 2 GHz to 15 GHz, 50 Ohms, Directivity: >16 dB,
Max VSWR: 1.35:1, Transmission Arm Loss: <1.5 dB, Coupled Arm Loss:
~10 dB,; Agilent 87300B or equivalent
Filter, Low Pass, 300 MHz, 50 Ohms, Rejection at >435 MHz: >45 dB;
Agilent 0955-0455 or equivalent
Filter, Low Pass, 1 GHz, 50 Ohms, Rejection at >2 GHz: >60 dB;
Agilent 0955-0487 or equivalent
Filter, Low Pass, 1.8 GHz, 50 Ohms, Rejection at 3 GHz: >45 dB,
2 ea.; Agilent 0955-0491 or equivalent
Filter, Low Pass, 4.4 GHz, 50 Ohms, Rejection at 5.5 GHz: >50 dB,
2 ea.; Agilent 11689A or equivalent
The average run times during testing were:
2 hr., 24 min. (TAR mode)