Package: agilent_e4405b_1yr_ver_9640_4418_hfsg_lfctr_lo_swpg.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
AT E4405B: (1 yr) VER /9640,4418,HFSG,LFCTR,LO,SWPG, Revision: 1.1
HP E4405B: (1 yr) VER /9640,4418,HFSG,LFCTR,LO,SWPG, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
AT E4405B: (1 yr) VER /9640,4418,HFSG,LFCTR,LO,SWPG
HP E4405B: (1 yr) VER /9640,4418,HFSG,LFCTR,LO,SWPG
Date: 2008-04-01
Revision: 1.1
Automated Reference: Agilent E4418B
Automated Reference: Fluke 9640A
Automated Reference: Fluke 9640A-50
Automated Reference: HF Signal Generator
Automated Reference: HP 8481A
Automated Reference: LF Counter
Automated Reference: Local Oscillator
Automated Reference: Sweep Generator
Manual Reference: Agilent 11667A
Manual Reference: Agilent 86205A
Manual Reference: Agilent 87300B
Model(s) Covered:
E4405B
Source(s):
Calibration Guide, Agilent Technologies ESA Spectrum Analyzers,
Manufacturing Part No. E4401-90475, March 2005
USAF T.O. 33K3-4-3314-1, Technical Manual, Calibration Procedure
For Spectrum Analyzer, E4405B, E4405B OPT AYX, 1DN, 1DR, 1D5, dated
30 April 2004
Compatibility:
MET/CAL 7.20x or later
Flexible Standard(s) Required:
Synthesized Sweep Generator (SWPG):
Frequency Range: 10 MHz to 14 GHz, CW
Accuracy: ±0.02% (Internal); External Reference capable
Amplitude Range: -40 dBm to +16 dBm
Resolution: 0.05 dB
SSB Phase Noise: <-80 dBc/Hz @ 50 kHz offset
Example: HP 83630B or equivalent
HF Signal Generator (HFSG):
Frequency Range: 1 GHz to 2.5 GHz
Accuracy: ±10 kHz (Internal); External Reference capable
Amplitude Range: -35 to +16 dBm
Resolution: 0.05 dB
Phase Noise at 1 GHz:
<-120 dBc/Hz @ 20 kHz offset
<-131 dBc/Hz @ 100 kHz offset
<-145 dBc/Hz @ 1 MHz offset
<-147 dBc/Hz @ 5 MHz offset
<-149 dBc/Hz @ 10 MHz offset
Example: HP 8663A or equivalent
Local Oscillator (LO):
Frequency Range: 10 MHz to 14 GHz, CW
Accuracy: ±0.02% (Internal); External Reference capable
Amplitude Range: -40 dBm to +16 dBm
Accuracy: ±2 dB
SSB Phase Noise: <-80 dBc/Hz @ 50 kHz offset
Example: HP 8340() or equivalent
LF Counter (LFCTR):
Ch. A Frequency Range: 10 MHz
Accuracy: Based on UUT timebase, ±0.6/0.025 ppm
External Reference capable
Example: Fluke PM 6690/6xx or equivalent
Frequency Standard:
Frequency Range: Suitable for use with configured LF counter
Accuracy: ±0.025 ppm or better (Option 1D5)
±0.6 ppm or better (standard)
(Internal/External frequency reference as required)
Additional Equipment Required:
Power Splitter, 1 kHz to 18 GHz, 6 dB insertion loss, output
tracking: <0.25 dB, equivalent output SWR: <1.22, HP/Agilent
11667A or equivalent
4.4 GHz Low Pass Filter, Rejection at >5.5 GHz: > 50 dB, 2 ea.,
HP/Agilent 11689A or equivalent
1.8 GHz Low Pass Filter, Rejection at >3 GHz: > 45 dB, 2 ea.,
HP/Agilent 0955-0491 or equivalent
1 GHz Low Pass Filter, Rejection at >2 GHz: > 60 dB, HP/Agilent
0955-0487 or equivalent
300 MHz Low Pass Filter, Rejection at >435 MHz: > 45 dB, HP/Agilent
0955-0455 or equivalent
Directional Coupler, 2 GHz to 15 GHz, Directivity: >16 dB, Max VSWR:
1.35:1, Transmission Arm Loss: <1.5 dB, Coupled Arm Loss: ~10 dB,
HP/Agilent 87300B or equivalent
Directional Bridge, 5 MHz to 3 GHz, Directivity: >40 dB, Coupling
Factor: 16 dB, Insertion Loss: 2 dB max, HP/Agilent 86205A or
equivalent
The average run times during testing were:
1 hr., 58 min. (TAR mode, Option 1DR equipped)