AT E4407B: (1 yr) VER /9640-LPN,4418,LFCTR,SWPG,UWSG

HP E4407B: (1 yr) VER /9640-LPN,4418,LFCTR,SWPG,UWSG

Date: 2012-12-06 17:00:53

Revision: 1.4

Automated Reference: Agilent E4418B

Automated Reference: Agilent E4419B

Automated Reference: Fluke 9640A

Automated Reference: Fluke 9640A-50

Automated Reference: HP 8481A

Automated Reference: HP 8481D

Automated Reference: HP 8485A

Automated Reference: Low Frequency Counter

Automated Reference: Microwave Sig Gen

Automated Reference: Sweep Generator

Manual Reference: Agilent 11667B

Manual Reference: Agilent 86205A

Manual Reference: Agilent 87300B

Model(s) Covered:

E4407B

Source(s):

Calibration Guide, Agilent Technologies ESA Spectrum Analyzers,

Manufacturing Part No. E4401-90493, March 2005

Compatibility:

MET/CAL 7.20 SP3 (7.20Z) or later

Flexible Standard(s) Required:

Synthesized Sweep Generator (SWPG):

Frequency Range: 10 MHz to 26.5 GHz, CW

Frequency Accuracy: ±0.02% (Internal); External Ref capable

Amplitude Range: -40 dBm to +16 dBm

Amplitude Resolution: 0.05 dB

SSB Phase Noise: <-80 dBc/Hz @ 50 kHz offset

Example: HP 83630B or equivalent

µW Signal Generator (UWSG):

Frequency Range: 10 MHz to 26.5 GHz, CW

Frequency Accuracy: ±0.02% (Internal); External Ref capable

Amplitude Range: -40 dBm to +16 dBm

Amplitude Accuracy: ±2 dB

SSB Phase Noise: <-80 dBc/Hz @ 50 kHz offset

Example: Anritsu MG3694() or equivalent

Low Frequency Counter (LFCTR):

Ch. A Frequency Range: 10 MHz

Ch. A Frequency Accuracy: Based on external frequency standard

Example: Fluke PM 6690/6xx or equivalent

Frequency Standard:

Frequency Range: Suitable for use with configured LF counter

Frequency Accuracy: ±0.025 ppm or better (Option 1D5)

±0.625 ppm or better (standard)

Additional Equipment Required:

Splitter,Power, 10 MHz to 26.5 GHz, 50 Ohms, 3.5 mm; Agilent® 11667B

or equivalent

Filter, Low Pass, 300 MHz, 50 Ohms, Rejection at >435 MHz: >45 dB;

Agilent 0955-0455 or equivalent

Filter, Low Pass, 1 GHz, 50 Ohms, Rejection at >2 GHz: >60 dB;

Agilent 0955-0487 or equivalent

Filter, Low Pass, 1.8 GHz, 50 Ohms, Rejection at 3 GHz: >45 dB,

2 ea.; Agilent 0955-0491 or equivalent

Filter, Low Pass, 4.4 GHz, 50 Ohms, Rejection at 5.5 GHz: >50 dB,

2 ea.; Agilent 11689A or equivalent

Adapter, Type N (m) - 3.5 (m), 50 Ohms, DC to 26.5 GHz, VSWR: 1.08;

Agilent 1250-1743 or equivalent

Coupler, Directional, 2 GHz to 15 GHz, 50 Ohms, Directivity: >16 dB,

Max VSWR: 1.35:1, Transmission Arm Loss: <1.5 dB, Coupled Arm Loss:

~10 dB,; Agilent 87300B or equivalent

Bridge, Directional, 5 MHz to 3 GHz, 50 Ohms, Directivity: >40 dB,

Coupling Factor: 16 dB, Insertion Loss: 2 dB max; Agilent 86205A

or equivalent

The average run times during testing were:

2 hr., 16 min. (TAR mode)