Package: agilent_e4407b_1yr_ver_9640_4418_hfsg_lfctr_lo_swpg.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

AT E4407B: (1 yr) VER /9640,4418,HFSG,LFCTR,LO,SWPG

HP E4407B: (1 yr) VER /9640,4418,HFSG,LFCTR,LO,SWPG

Date: 2008-04-01

Revision: 1.2

Automated Reference: Agilent E4418B

Automated Reference: Fluke 9640A

Automated Reference: Fluke 9640A-50

Automated Reference: HF Signal Generator

Automated Reference: HP 8481D

Automated Reference: HP 8485A

Automated Reference: LF Counter

Automated Reference: Local Oscillator

Automated Reference: Sweep Generator

Manual Reference: Agilent 11667B

Manual Reference: Agilent 86205A

Manual Reference: Agilent 87300B

Model(s) Covered:

E4407B

Source(s):

Calibration Guide, Agilent Technologies ESA Spectrum Analyzers,

Manufacturing Part No. E4401-90475, March 2005

USAF T.O. 33K3-4-3334-1, Technical Manual, Calibration Procedure

For Spectrum Analyzer, E4407B, dated 30 June 2002, Change 8,

28 February 2007

Compatibility:

MET/CAL 7.20x or later

Flexible Standard(s) Required:

Synthesized Sweep Generator (SWPG):

Frequency Range: 10 MHz to 26.5 GHz, CW

Accuracy: ±0.02% (Internal); External Reference capable

Amplitude Range: -40 dBm to +16 dBm

Resolution: 0.05 dB

SSB Phase Noise: <-80 dBc/Hz @ 50 kHz offset

Example: HP 83630B or equivalent

HF Signal Generator (HFSG):

Frequency Range: 1 GHz to 2.5 GHz

Accuracy: ±10 kHz (Internal); External Reference capable

Amplitude Range: -35 to +16 dBm

Resolution: 0.05 dB

Phase Noise at 1 GHz:

<-120 dBc/Hz @ 20 kHz offset

<-131 dBc/Hz @ 100 kHz offset

<-145 dBc/Hz @ 1 MHz offset

<-147 dBc/Hz @ 5 MHz offset

<-149 dBc/Hz @ 10 MHz offset

Example: HP 8663A or equivalent

Local Oscillator (LO):

Frequency Range: 10 MHz to 26.5 GHz, CW

Accuracy: ±0.02% (Internal); External Reference capable

Amplitude Range: -40 dBm to +16 dBm

Accuracy: ±2 dB

SSB Phase Noise: <-80 dBc/Hz @ 50 kHz offset

Example: HP 8340() or equivalent

LF Counter (LFCTR):

Ch. A Frequency Range: 10 MHz

Accuracy: Based on external frequency standard

Ch. C Frequency Range: 300 MHz

Accuracy: Based on external frequency standard

Time Interval Range: 25 ms to 100 ms

Accuracy: ±0.25% or better (internal frequency reference)

Example: Fluke PM 6690/6xx or equivalent

Frequency Standard:

Frequency Range: Suitable for use with configured LF counter

Accuracy: ±0.03 ppm or better (standard)

±0.625 ppm or better (Option 103)

Additional Equipment Required:

Power Splitter, 1 kHz to 26.5 GHz, 6 dB insertion loss, output

tracking: <0.25 dB, equivalent output SWR: <1.22, HP/Agilent

11667B or equivalent

4.4 GHz Low Pass Filter, Rejection at >5.5 GHz: > 50 dB, 2 ea.,

HP/Agilent 11689A or equivalent

1.8 GHz Low Pass Filter, Rejection at >3 GHz: > 45 dB, 2 ea.,

HP/Agilent 0955-0491 or equivalent

1 GHz Low Pass Filter, Rejection at >2 GHz: > 60 dB, HP/Agilent

0955-0487 or equivalent

300 MHz Low Pass Filter, Rejection at >435 MHz: > 45 dB, HP/Agilent

0955-0455 or equivalent

Directional Coupler, 2 GHz to 15 GHz, Directivity: >16 dB, Max VSWR:

1.35:1, Transmission Arm Loss: <1.5 dB, Coupled Arm Loss: ~10 dB,

HP/Agilent 87300B or equivalent

Directional Bridge, 5 MHz to 3 GHz, Directivity: >40 dB, Coupling

Factor: 16 dB, Insertion Loss: 2 dB max, HP/Agilent 86205A or

equivalent

The average run times during testing were:

1 hr., 58 min. (TAR mode, Option 1DR equipped)