AT E4408B: (1 yr) CAL VER /9640,4418,LFCTR,LO,SWPG, Revision: 1.1
HP E4408B: (1 yr) CAL VER /9640,4418,LFCTR,LO,SWPG, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
AT E4408B: (1 yr) CAL VER /9640,4418,LFCTR,LO,SWPG
HP E4408B: (1 yr) CAL VER /9640,4418,LFCTR,LO,SWPG
Date: 2008-04-01
Revision: 1.1
Automated Reference: Agilent E4418B
Automated Reference: Fluke 9640A
Automated Reference: Fluke 9640A-50
Automated Reference: HP 8485A
Automated Reference: LF Counter
Automated Reference: Local Oscillator
Automated Reference: Sweep Generator
Manual Reference: Agilent 11667B
Manual Reference: Agilent 86205A
Manual Reference: Agilent 87300B
Model(s) Covered:
E4408B
Source(s):
Calibration Guide, Agilent Technologies ESA Spectrum Analyzers,
Manufacturing Part No. E4401-90475, March 2005
USAF T.O. 33K3-4-3283-1, Technical Manual, Calibration Procedure
For Spectrum Analyzer, E4408B, dated 30 June 2001, Change 6,
30 June 2006
Compatibility:
MET/CAL 7.20x or later
Flexible Standard(s) Required:
Synthesized Sweep Generator (SWPG):
Frequency Range: 10 MHz to 26.5 GHz, CW
Accuracy: ±0.02% (Internal); External Reference capable
Amplitude Range: -40 dBm to +16 dBm
Resolution: 0.05 dB
SSB Phase Noise: <-80 dBc/Hz @ 50 kHz offset
Example: HP 83630B or equivalent
Local Oscillator (LO):
Frequency Range: 10 MHz to 26.5 GHz, CW
Accuracy: ±0.02% (Internal); External Reference capable
Amplitude Range: -40 dBm to +16 dBm
Accuracy: ±2 dB
SSB Phase Noise: <-80 dBc/Hz @ 50 kHz offset
Example: HP 8340() or equivalent
LF Counter (LFCTR):
Ch. A Frequency Range: 10 MHz
Accuracy: Based on external frequency standard
Ch. C Frequency Range: 300 MHz
Accuracy: Based on external frequency standard
Time Interval Range: 25 ms to 100 ms
Accuracy: ±0.25% or better (internal frequency reference)
Example: Fluke PM 6690/6xx or equivalent
Frequency Standard:
Frequency Range: Suitable for use with configured LF counter
Accuracy: ±0.03 ppm or better (standard)
±0.625 ppm or better (Option 103)
Additional Equipment Required:
Power Splitter, 1 kHz to 26.5 GHz, 6 dB insertion loss, output
tracking: <0.25 dB, equivalent output SWR: <1.22, HP/Agilent
11667B or equivalent
4.4 GHz Low Pass Filter, Rejection at >5.5 GHz: > 50 dB, 2 ea.,
HP/Agilent 11689A or equivalent
1.8 GHz Low Pass Filter, Rejection at >3 GHz: > 45 dB, 2 ea.,
HP/Agilent 0955-0491 or equivalent
1 GHz Low Pass Filter, Rejection at >2 GHz: > 60 dB, HP/Agilent
0955-0487 or equivalent
300 MHz Low Pass Filter, Rejection at >435 MHz: > 45 dB, HP/Agilent
0955-0455 or equivalent
Directional Coupler, 2 GHz to 15 GHz, Directivity: >16 dB, Max VSWR:
1.35:1, Transmission Arm Loss: <1.5 dB, Coupled Arm Loss: ~10 dB,
HP/Agilent 87300B or equivalent
Directional Bridge, 5 MHz to 3 GHz, Directivity: >40 dB, Coupling
Factor: 16 dB, Insertion Loss: 2 dB max, HP/Agilent 86205A or
equivalent
The average run times during testing were:
1 hr., 50 min.