Package: agilent_e4411b_1yr_ver_9640_4418_hfsg_lfctr.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
AT E4411B: (1 yr) CAL VER /9640,4418,HFSG,LFCTR, Revision: 1.4
HP E4411B: (1 yr) CAL VER /9640,4418,HFSG,LFCTR, Revision: 1.4
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
AT E4411B: (1 yr) CAL VER /9640,4418,HFSG,LFCTR
HP E4411B: (1 yr) CAL VER /9640,4418,HFSG,LFCTR
Date: 2011-09-20
Revision: 1.4
Automated Reference: Agilent E4418B
Automated Reference: Fluke 9640A
Automated Reference: Fluke 9640A-50
Automated Reference: HF Signal Generator
Automated Reference: HP 8482A
Automated Reference: LF Counter
Automated Reference: Low Frequency Counter
Manual Reference: Agilent 86205A
Model(s) Covered:
E4411B
Source(s):
Calibration Guide, Agilent Technologies ESA Spectrum Analyzers,
Manufacturing Part No. E4401-90493, March 2005
Compatibility:
MET/CAL 7.20 SP3 (7.20Z) or later
Flexible Standard(s) Required:
HF Signal Generator (HFSG):
Frequency Range: 50 MHz to 1.5 GHz, CW
Accuracy: ±0.02% (Internal); External Reference capable
Amplitude Range: -40 dBm to +16 dBm
Accuracy: ±2 dB
SSB Phase Noise: <-80 dBc/Hz @ 50 kHz offset
Example: Anritsu MG3690() or equivalent
LF Counter (LFCTR):
Ch. A Frequency Range: 10 MHz
Accuracy: Based on external frequency standard
Example: Fluke PM 6690/6xx or equivalent
Frequency Standard:
Frequency Range: Suitable for use with configured LF counter
Accuracy: ±0.625 ppm or better
Additional Equipment Required:
Bridge, Directional, 5 MHz to 3 GHz, 50 Ohms, Directivity: >40 dB,
Coupling Factor: 16 dB, Insertion Loss: 2 dB max; Agilent 86205A
or equivalent
Filter, Low Pass, 50 MHz, Rejection >65 MHz: >40 dB, Rejection
>75 MHz: >60 dB; Agilent 0955-0306 or equivalent
The average run times during testing were:
39 min. (TAR mode)