AT E4411B: (1 yr) CAL VER /9640,4418,HFSG,LFCTR

HP E4411B: (1 yr) CAL VER /9640,4418,HFSG,LFCTR

Date: 2011-09-20

Revision: 1.4

Automated Reference: Agilent E4418B

Automated Reference: Fluke 9640A

Automated Reference: Fluke 9640A-50

Automated Reference: HF Signal Generator

Automated Reference: HP 8482A

Automated Reference: LF Counter

Automated Reference: Low Frequency Counter

Manual Reference: Agilent 86205A

Model(s) Covered:

E4411B

Source(s):

Calibration Guide, Agilent Technologies ESA Spectrum Analyzers,

Manufacturing Part No. E4401-90493, March 2005

Compatibility:

MET/CAL 7.20 SP3 (7.20Z) or later

Flexible Standard(s) Required:

HF Signal Generator (HFSG):

Frequency Range: 50 MHz to 1.5 GHz, CW

Accuracy: ±0.02% (Internal); External Reference capable

Amplitude Range: -40 dBm to +16 dBm

Accuracy: ±2 dB

SSB Phase Noise: <-80 dBc/Hz @ 50 kHz offset

Example: Anritsu MG3690() or equivalent

LF Counter (LFCTR):

Ch. A Frequency Range: 10 MHz

Accuracy: Based on external frequency standard

Example: Fluke PM 6690/6xx or equivalent

Frequency Standard:

Frequency Range: Suitable for use with configured LF counter

Accuracy: ±0.625 ppm or better

Additional Equipment Required:

Bridge, Directional, 5 MHz to 3 GHz, 50 Ohms, Directivity: >40 dB,

Coupling Factor: 16 dB, Insertion Loss: 2 dB max; Agilent 86205A

or equivalent

Filter, Low Pass, 50 MHz, Rejection >65 MHz: >40 dB, Rejection

>75 MHz: >60 dB; Agilent 0955-0306 or equivalent

The average run times during testing were:

39 min. (TAR mode)