Agilent U1242A: (1 year) CAL VER /5500

Date: 2013-07-10 14:54:43

Revision: 2619

Automated Reference: Fluke 5500A

Model(s) Covered:

U1242A

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5500A (1 year) specifications.

RESISTANCE TESTS

100 MOhm Range: 100.00 MOhm; TUR 3.7731

CAPACITANCE TESTS

10 µF Range: 10.000 µF; TUR 3.4444

100 µF Range: 100.00 µF; TUR 2.5833

1000 µF Range: 1000.0 µF; TUR 1.9806

10 mF Range: 1.100 mF; TUR 2.3009

Note: The 10 mF Range test point deviates from the manual,

testing at 1.1 mF instead of 10 mF, due to the 5500A's

Synthesized Capacitance capability.

Source:

Agilent U1241A and U1242A Handheld Digital Multimeters

User's and Service Guide

PN: U1241-90003, Seventh Edition, May 4, 2012

Compatibility:

MET/CAL version 7.3.38 or later

Additional Equipment Required:

Fluke 80AK Type-K DMM (thermocouple) Adapter

Type-K thermocouple extension cable with SMP (m)-(m) plugs

The average run time during testing was 15 minutes (NMEAS = 1).