Agilent U1241A: (1 year) CAL VER /5500

Agilent U1242A: (1 year) CAL VER /5500

Date: 2009-08-11 11:55:20

Revision: 1.1

Automated Reference: Fluke 5500A

Model(s) Covered:

U1241A, U1242A

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5500A (1 year) specifications.

RESISTANCE TESTS

100 MOhm Range: 100.00 MOhm; TUR 3.0267

CAPACITANCE TESTS

1000 nF Range: 1000.0 nF; TUR 3.5429

10 µF Range: 10.000 µF; TUR 2.7556

100 µF Range: 100.00 µF; TUR 2.0677

1000 µF Range: 1000.0 µF; TUR 1.9806

10 mF Range: 1.100 mF; TUR 2.3009

Source:

Agilent U1241A and U1242A Handheld Digital Multimeters

User's and Service Guide

PN: U1241-90003, Second Edition, April 01, 2009

Compatibility:

MET/CAL or 5500/CAL 7.20C or later

Additional Equipment Required:

Fluke 80AK Type-K DMM (thermocouple) Adapter

Type-K thermocouple extension cable with SMP (m)-(m) plugs

The average run time during testing was 15 minutes (keyboard entry).