Package: agilent_u124xa_cal_5520.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Agilent U1241A: (1 year) CAL VER /5520, Revision: 1.1
Agilent U1242A: (1 year) CAL VER /5520, Revision: 1.1
Agilent U1241A: CAL ADJ /5520, Revision: 1.2
Agilent U1242A: CAL ADJ /5520, Revision: 1.2
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Agilent U1241A: (1 year) CAL VER /5520
Agilent U1242A: (1 year) CAL VER /5520
Date: 2009-08-12 08:13:54
Revision: 1.1
Automated Reference: Fluke 5520A
Date: 2009-08-12 08:13:56
Revision: 1.2
Automated Reference: Fluke 5520A
Model(s) Covered:
U1241A, U1242A
Source:
Agilent U1241A and U1242A Handheld Digital Multimeters
User's and Service Guide
PN: U1241-90003, Second Edition, April 01, 2009
Compatibility:
MET/CAL or 5500/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
Fluke 80AK Type-K DMM (thermocouple) Adapter
Type-K thermocouple extension cable with SMP (m)-(m) plugs
The average run time during testing was 20 minutes.
Model(s) Covered:
U1241A, U1242A
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 5520A (1 year) specifications.
CAPACITANCE TESTS
1000 nF Range: 1000.0 nF; TUR 3.5429
10 µF Range: 10.000 µF; TUR 3.5429
100 µF Range: 100.00 µF; TUR 2.2545
1000 µF Range: 1000.0 µF; TUR 3.7091
10 mF Range: 10.000 mF; TUR 3.7091
Source:
Agilent U1241A and U1242A Handheld Digital Multimeters
User's and Service Guide
PN: U1241-90003, Second Edition, April 01, 2009
Compatibility:
MET/CAL or 5500/CAL 7.20C or later
Additional Equipment Required:
Fluke 80AK Type-K DMM (thermocouple) Adapter
Type-K thermocouple extension cable with SMP (m)-(m) plugs
The average run time during testing was 15 minutes (keyboard entry).