Package: agilent_u124xa_cal_5520.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Agilent U1241A: (1 year) CAL VER /5520

Agilent U1242A: (1 year) CAL VER /5520

Date: 2009-08-12 08:13:54

Revision: 1.1

Automated Reference: Fluke 5520A

Model(s) Covered:

U1241A, U1242A

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5520A (1 year) specifications.

CAPACITANCE TESTS

1000 nF Range: 1000.0 nF; TUR 3.5429

10 µF Range: 10.000 µF; TUR 3.5429

100 µF Range: 100.00 µF; TUR 2.2545

1000 µF Range: 1000.0 µF; TUR 3.7091

10 mF Range: 10.000 mF; TUR 3.7091

Source:

Agilent U1241A and U1242A Handheld Digital Multimeters

User's and Service Guide

PN: U1241-90003, Second Edition, April 01, 2009

Compatibility:

MET/CAL or 5500/CAL 7.20C or later

Additional Equipment Required:

Fluke 80AK Type-K DMM (thermocouple) Adapter

Type-K thermocouple extension cable with SMP (m)-(m) plugs

The average run time during testing was 15 minutes (keyboard entry).

Agilent U1241A: CAL ADJ /5520

Agilent U1242A: CAL ADJ /5520

Date: 2009-08-12 08:13:56

Revision: 1.2

Automated Reference: Fluke 5520A

Model(s) Covered:

U1241A, U1242A

Source:

Agilent U1241A and U1242A Handheld Digital Multimeters

User's and Service Guide

PN: U1241-90003, Second Edition, April 01, 2009

Compatibility:

MET/CAL or 5500/CAL 7.20 SP1 (7.20x) or later

Additional Equipment Required:

Fluke 80AK Type-K DMM (thermocouple) Adapter

Type-K thermocouple extension cable with SMP (m)-(m) plugs

The average run time during testing was 20 minutes.