Agilent U3402A: (1 year) CAL VER /5520, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Agilent U3402A: (1 year) CAL VER /5520
Date: 2011-10-12 06:36:36
Revision: 1.1
Automated Reference: Fluke 5520A
Model(s) Covered:
U3402A
Source:
Agilent U3402A 5 1/2 - Digit Dual Display Multimeter
User's and Service Guide
Compatibility:
MET/CAL or 5500/CAL 7.3.38 or later
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1 based on 5520A 1 year specifications.
DC CURRENT TESTS:
12 A Range (Slow)
10 A; TUR 3.7273
Additional Equipment Required:
None
The average run time during testing was 37 minutes (TAR mode).
The average run time during testing was <tbd> minutes (MU mode).