Package: amprobe_38xr_1yr_ver_rs232_5520.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Amprobe 38XR-A: (1 year) CAL VER /5520

Meterman 38XR: (1 year) CAL VER /5520

Date: 2007-04-16 15:21:02

Revision: 1.4

Automated Reference: Fluke 5520A

Manual Reference: GenRad 1409-L

Manual Reference: GenRad 1409-T

Manual Reference: GenRad 1409-Y

Manual Reference: IET 1409-100uF

Manual Reference: IET 1409-10uF

Model(s) Covered:

Meterman 38XR

Amprobe 38XR-A

Source:

Meterman 38XR Operating Manual, PN 1990734, Copyright Oct 2002

Compatibility:

MET/CAL or 5500/CAL 7.20 or later

Additional Equipment Required:

None

The average run time during testing was 20 minutes (keyboard entry).

Amprobe 38XR-A: (1 year) CAL VER RS-232 /5520

Meterman 38XR: (1 year) CAL VER RS-232 /5520

Date: 2007-04-16 15:21:20

Revision: 1.2

Automated Reference: Fluke 5520A

Manual Reference: GenRad 1409-L

Manual Reference: GenRad 1409-T

Manual Reference: GenRad 1409-Y

Manual Reference: IET 1409-100uF

Manual Reference: IET 1409-10uF

Model(s) Covered:

Meterman 38XR

Amprobe 38XR-A

Source:

Meterman 38XR Professional Digital Multimeter, Users Manual,

Part No. 1990734, October 2002

Compatibility:

MET/CAL or 5500/CAL 7.20 or later

Additional Equipment Required:

- Meterman 38SW SERIAL CABLE

The average run time during testing was 15 minutes.