Package: at_34410a_1yr_cal_ver_ieee_5720_fgen.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Agilent 34410A: (1 year) CAL VER IEEE /5720,FGEN, Revision: 2
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Agilent 34410A: (1 year) CAL VER IEEE /5720,FGEN
Date: 2011-08-18 14:39:00
Revision: 2
Automated Reference: Fluke 5720A
Automated Reference: Function Generator
Model Covered:
34410A
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1 using 5720A 1 year specifications.
AC VOLTAGE
- 1 V, 10 Hz; TUR 2.9062
- 10 V, 10 Hz; TUR 2.8219
AC CURRENT
- 100 µA, 5 kHz; TUR 1.2727
- 1 mA, 10 kHz; TUR 1.2632
- 10 mA, 10 kHz; TUR 1.4118
- 100 mA, 10 kHz; TUR 1.9672
- 1 A, 5 kHz; TUR 2.5000
Source:
Agilent 34410A/11A 6½ Digit Multimeter Service Guide, Part No.
34410-90010, Third Edition, April 2006
Compatibility:
MET/CAL 7.3 SP1 (7.3.36) or later
Additional Equipment Required:
Precision 4-wire short with standard ¾" banana spacing, Agilent
Model 34172A DMM Calibration Short or equivalent
Cable, 50 Ohm, BNC(m)-(m), Pomona 2249-C-36
BNC female to dual banana plug, Pomona 1269
The average run time during testing was 23 minutes (TAR).