Package: at_34970a_1yr_cal_ver_ieee_5700_fgen.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Agilent 34970A: (1 year) CAL VER IEEE /5700,FGEN, Revision: 1.2
HP 34970A: (1 year) CAL VER IEEE /5700,FGEN, Revision: 1.2
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Agilent 34970A: (1 year) CAL VER IEEE /5700,FGEN
HP 34970A: (1 year) CAL VER IEEE /5700,FGEN
Date: 2011-08-17 14:51:41
Revision: 1.2
Automated Reference: Fluke 5700A
Automated Reference: Function Generator
Models covered:
HP/Agilent 34970A w/34901A 20 Channel Multiplexer module
The following test points have a Test Uncertainty Ratio (TUR) of less than
4 to 1 based 5700A 1 year specifications.
AC VOLTAGE, 10 V Range : 100 mV @ 50 kHz : TUR 3.9535
AC VOLTAGE, 10 V Range : 10 V @ 10 Hz : TUR 1.5385
OHMS 2 WIRE, 100 Ohm Range : 100 Ohms : TUR 3.7838
AC CURRENT, 1 A Range : 1 A @ 1 kHz : TUR 2.0290
Compatibility:
MET/CAL 7.3 SP1 (7.3.36) or later
Required Equipment:
HP/Agilent 34901A 20 Channel Multiplexer module
Copper wire, insulated, ~15 cm long
Leads, Teflon®, twisted pair, shielded, 3 ea. sufficiently long to
connect UUT to source
50 Ohm BNC(m)-(m) cable, 1 set
Adapter, BNC(f) to binding posts, Pomona 1452
50 Ohm feed through terminator, 1 set
(Only for generator having the fixed 50 Ohm impedance)
The average run time during testing was:
15 minutes (Root Difference Square guardbanding method Mode (TAR))
28 minutes (Measurement Uncertainty guardbanding method Mode (MU))