Package: at_dso9254a_1yr_ver_visa_9500_4hd_lo.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Agilent DSO9254A: (1 yr) CAL VER VISA /9500+4Hd,LO

Date: 2012-01-18 10:24:31

Revision: 1.3

Automated Reference: Datron 9500

Automated Reference: Datron 9510

Automated Reference: Datron 9530

Automated Reference: Local Oscillator

Models covered:

DSO9254A

Source(s):

Agilent 9000 Series Infiniium Oscilloscopes, Service Guide

PN 54904-97005, October 2009

Compatibility:

MET/CAL 7.3 or later

Flexible Standard(s) Required:

- Local Oscillator (LO):

Example: Agilent E8257D Opt 520 or equivalent

- Frequency Range: 10.00002 MHz, CW

Accuracy: < ± 0.01 ppm

Settability: 10 Hz (Internal)

- Amplitude Range: 0 dBm (into 50 Ohms)

Additional Equipment Required:

- Adapter, SMA(f) to BNC(m)

only for 9560 active head

- Output cable and necessary adapter to BNC(m)

depends on Local Oscillator (LO) RF Output type