Package: avetek_2015_1yr_ver_9100.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Wavetek 2015: (1 year) CAL VER /9100

Date: 2005-02-08

Revision: 1.1

Automated Reference: Datron 9100

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 9100 (1 year) specifications.

Capacitance Tests:

4 nF Range, 3.500nF; TUR 2.9412

40 nF Range, 35.00nF; TUR 2.8889

400 nF Range, 350.0nF; TUR 3.2231

4 µF Range, 3.500uF; TUR 2.5000

40 µF Range, 10.00uF; TUR TUR 2.1212

Model(s) Covered:

Wavetek 2015

Source:

Wavetek 2015 Operator's Manual, part number 1566149

Compatibility:

MET/CAL 7.01 SP2 or later

Additional Equipment Required:

None

The average run time during testing was 15 minutes (keyboard entry).