Package: benning_multitester_iii_1yr_ver_9100.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

BENNING MULTITESTER III: (1 yr) CAL VER /9100

Date: 2007-03-23 13:47:42

Revision: 1.1

Automated Reference: Datron 9100

Model(s) Covered: Multitester III

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1.

CAPACITANCE

1.9 nF; TUR 2.0290

19 µF; TUR 3.7838

190 µF; TUR 3.7838

Source: BENNING MULTITESTER III Operating Manual, 1/99

Section 7 Electrical Specifications

Compatibility: MET/CAL 7.01 or later

Additional Equipment Required: None

An special cable is required to connect the 9100 Hi & Lo

terminals to the UUT + and - capacitance terminals.

The average run time during testing was 15 minutes.