Clarke-Hess 6000: (1 year) CAL VER IEEE /CH5500,DMM
Date: 2010-07-01 11:41:05
Revision: 1.2
Automated Reference: Clarke-Hess 5500
Automated Reference: Digital Multimeter
Model(s) Covered:
Clarke-Hess 6000
This procedure contains tests whose Test Uncertainty Ratios are less than
4:1.
ANGLE TRACKING ACCURACY TESTS
-160°, 160 mV @ 50 kHz; TUR 3.3333
-150°, 160 mV @ 50 kHz; TUR 3.3333
-140°, 160 mV @ 50 kHz; TUR 3.3333
-130°, 160 mV @ 50 kHz; TUR 3.3333
-120°, 160 mV @ 50 kHz; TUR 3.3333
-110°, 160 mV @ 50 kHz; TUR 3.3333
-100°, 160 mV @ 50 kHz; TUR 3.3333
-90°, 160 mV @ 50 kHz; TUR 3.3333
-80°, 160 mV @ 50 kHz; TUR 3.3333
-70°, 160 mV @ 50 kHz; TUR 3.3333
-60°, 160 mV @ 50 kHz; TUR 3.3333
-50°, 160 mV @ 50 kHz; TUR 3.3333
-40°, 160 mV @ 50 kHz; TUR 3.3333
-30°, 160 mV @ 50 kHz; TUR 3.3333
-20°, 160 mV @ 50 kHz; TUR 3.3333
-10°, 160 mV @ 50 kHz; TUR 3.3333
0°, 160 mV @ 50 kHz; TUR 3.3333
10°, 160 mV @ 50 kHz; TUR 3.3333
20°, 160 mV @ 50 kHz; TUR 3.3333
30°, 160 mV @ 50 kHz; TUR 3.3333
40°, 160 mV @ 50 kHz; TUR 3.3333
50°, 160 mV @ 50 kHz; TUR 3.3333
60°, 160 mV @ 50 kHz; TUR 3.3333
70°, 160 mV @ 50 kHz; TUR 3.3333
80°, 160 mV @ 50 kHz; TUR 3.3333
90°, 160 mV @ 50 kHz; TUR 3.3333
100°, 160 mV @ 50 kHz; TUR 3.3333
110°, 160 mV @ 50 kHz; TUR 3.3333
120°, 160 mV @ 50 kHz; TUR 3.3333
130°, 160 mV @ 50 kHz; TUR 3.3333
140°, 160 mV @ 50 kHz; TUR 3.3333
150°, 160 mV @ 50 kHz; TUR 3.3333
160°, 160 mV @ 50 kHz; TUR 3.3333
170°, 160 mV @ 50 kHz; TUR 3.3333
180°, 160 mV @ 50 kHz; TUR 3.3333
190°, 160 mV @ 50 kHz; TUR 3.3333
200°, 160 mV @ 50 kHz; TUR 3.3333
210°, 160 mV @ 50 kHz; TUR 3.3333
220°, 160 mV @ 50 kHz; TUR 3.3333
230°, 160 mV @ 50 kHz; TUR 3.3333
240°, 160 mV @ 50 kHz; TUR 3.3333
250°, 160 mV @ 50 kHz; TUR 3.3333
260°, 160 mV @ 50 kHz; TUR 3.3333
270°, 160 mV @ 50 kHz; TUR 3.3333
280°, 160 mV @ 50 kHz; TUR 3.3333
290°, 160 mV @ 50 kHz; TUR 3.3333
300°, 160 mV @ 50 kHz; TUR 3.3333
310°, 160 mV @ 50 kHz; TUR 3.3333
320°, 160 mV @ 50 kHz; TUR 3.3333
330°, 160 mV @ 50 kHz; TUR 3.3333
340°, 160 mV @ 50 kHz; TUR 3.3333
AMPLITUDE LINEARITY TESTS ("SIGNAL IN" constant)
60°, 600 mV @ 20 kHz; TUR 3.3333
60°, 800 mV @ 20 kHz; TUR 3.3333
60°, 1 V @ 20 kHz; TUR 3.3333
60°, 2 V @ 20 kHz; TUR 3.3333
60°, 4 V @ 20 kHz; TUR 3.3333
60°, 6 V @ 20 kHz; TUR 3.3333
60°, 8 V @ 20 kHz; TUR 3.3333
60°, 10 V @ 20 kHz; TUR 3.3333
60°, 12 V @ 20 kHz; TUR 3.3333
60°, 600 mV @ 50 kHz; TUR 3.3333
60°, 800 mV @ 50 kHz; TUR 3.3333
60°, 1 V @ 50 kHz; TUR 3.3333
60°, 2 V @ 50 kHz; TUR 3.3333
60°, 4 V @ 50 kHz; TUR 3.3333
60°, 6 V @ 50 kHz; TUR 3.3333
60°, 8 V @ 50 kHz; TUR 3.3333
60°, 10 V @ 50 kHz; TUR 3.3333
60°, 12 V @ 50 kHz; TUR 3.3333
60°, 600 mV @ 100 kHz; TUR 3.75
60°, 800 mV @ 100 kHz; TUR 3.75
60°, 1 V @ 100 kHz; TUR 3.75
60°, 2 V @ 100 kHz; TUR 3.75
60°, 4 V @ 100 kHz; TUR 3.75
60°, 6 V @ 100 kHz; TUR 3.75
60°, 8 V @ 100 kHz; TUR 3.75
60°, 10 V @ 100 kHz; TUR 3.75
60°, 12 V @ 100 kHz; TUR 3.75
AMPLITUDE LINEARITY TESTS ("REFERENCE IN" constant)
60°, 600 mV @ 20 kHz; TUR 3.3333
60°, 800 mV @ 20 kHz; TUR 3.3333
60°, 1 V @ 20 kHz; TUR 3.3333
60°, 2 V @ 20 kHz; TUR 3.3333
60°, 4 V @ 20 kHz; TUR 3.3333
60°, 6 V @ 20 kHz; TUR 3.3333
60°, 8 V @ 20 kHz; TUR 3.3333
60°, 10 V @ 20 kHz; TUR 3.3333
60°, 12 V @ 20 kHz; TUR 3.3333
60°, 600 mV @ 50 kHz; TUR 3.3333
60°, 800 mV @ 50 kHz; TUR 3.3333
60°, 1 V @ 50 kHz; TUR 3.3333
60°, 2 V @ 50 kHz; TUR 3.3333
60°, 4 V @ 50 kHz; TUR 3.3333
60°, 6 V @ 50 kHz; TUR 3.3333
60°, 8 V @ 50 kHz; TUR 3.3333
60°, 10 V @ 50 kHz; TUR 3.3333
60°, 12 V @ 50 kHz; TUR 3.3333
60°, 600 mV @ 100 kHz; TUR 3.75
60°, 800 mV @ 100 kHz; TUR 3.75
60°, 1 V @ 100 kHz; TUR 3.75
60°, 2 V @ 100 kHz; TUR 3.75
60°, 4 V @ 100 kHz; TUR 3.75
60°, 6 V @ 100 kHz; TUR 3.75
60°, 8 V @ 100 kHz; TUR 3.75
60°, 10 V @ 100 kHz; TUR 3.75
60°, 12 V @ 100 kHz; TUR 3.75
AMPLITUDE RANGE TRACKING TESTS
Reference Signal: 160 mV @ 50 kHz
60°, 160 mV @ 50 kHz; TUR 3.3333 (Signal In)
60°, 4 V @ 50 kHz; TUR 3.3333 (Signal In)
Reference Signal: 4 V @ 50 kHz
60°, 160 mV @ 50 kHz; TUR 3.3333 (Signal In)
60°, 4 V @ 50 kHz; TUR 3.3333 (Signal In)
60°, 100 V @ 50 kHz; TUR 3.3333 (Signal In)
Reference Signal: 100 V @ 50 kHz
60°, 4 V @ 50 kHz; TUR 3.3333 (Signal In)
60°, 100 V @ 50 kHz; TUR 3.3333 (Signal In)
Reference Signal: 160 mV @ 100 kHz
60°, 160 mV @ 100 kHz; TUR 3.75 (Signal In)
60°, 4 V @ 100 kHz; TUR 3.75 (Signal In)
Reference Signal: 4 V @ 100 kHz
60°, 160 mV @ 100 kHz; TUR 3.75 (Signal In)
60°, 4 V @ 100 kHz; TUR 3.75 (Signal In)
60°, 100 V @ 100 kHz; TUR 3.75 (Signal In)
Reference Signal: 100 V @ 100 kHz
60°, 4 V @ 100 kHz; TUR 3.75 (Signal In)
60°, 100 V @ 100 kHz; TUR 3.75 (Signal In)
Source:
Clarke-Hess 6000 Instruction Manual, Section III, Performance Tests
Compatibility:
MET/CAL 7.3 or later
Flexible Standard(s) Required:
Digital Multimeter (DMM):
DC Voltage Range: ±4 V
Accuracy: ±0.1 % or less
Resolution: 100 µV or better
Example: Fluke 8845A or equivalent
Additional Equipment Required:
- Precision coaxial cable, 50 Ohm, BNC(m)-(m), 36", Tektronix
012-0482-00 or equivalent, 2 ea.
- Coaxial cable, 50 Ohm, BNC(m)-(m), 36", Pomona Model 2249-C-36
- Adapter, BNC(f) to double-banana plugs, Pomona Model 1269
The average run time during testing was 1 hour, 30 minutes.