CLIPSAL 486D: (1 year) CAL VER /5320,SCPI DMM

KYORITSU 5402D: (1 year) CAL VER /5320,SCPI DMM

Date: 2009-05-01 10:18:35

Revision: 1.5

Automated Reference: Digital Multimeter

Automated Reference: Fluke 5320A

Model Covered:

486D, 5402D

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on the 5320A and DMM 1 year specifications. The actual

TUR value may vary slightly for it depends on the DMM reading.

NO TRIP (½ Rated mA) CURRENT ACCURACY TESTS

2.50 mA @ 50 Hz : TUR 2.53

5.00 mA @ 50 Hz : TUR 2.99

15.0 mA @ 50 Hz : TUR 2.98

50.0 mA @ 50 Hz : TUR 3.08

150 mA @ 50 Hz : TUR 2.95

250 mA @ 50 Hz : TUR 2.97

TRIP (Rated mA) CURRENT ACCURACY TESTS

5.00 mA @ 50 Hz : TUR 3.01

10.0 mA @ 50 Hz : TUR 2.99

30.0 mA @ 50 Hz : TUR 3.02

100 mA @ 50 Hz : TUR 3.04

300 mA @ 50 Hz : TUR 3.02

500 mA @ 50 Hz : TUR 3.05

FAST TRIP (250 mA) CURRENT ACCURACY TESTS

250 mA @ 50 Hz : TUR 2.98