Package: escort_178_1yr_ver_9100.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Escort 178: (1 year) CAL VER /9100

Date: 2010-06-23 15:24:37

Revision: 1.2

Automated Reference: Datron 9100

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 9100 (1 year) specifications.

DC CURRENT TESTS:

400 mA Range, 390.0mA; TUR 3.0682

40 mA Range, -390.0mA; TUR 3.0682

CAPACITANCE TESTS:

40 µF Range, 39.00µF; TUR 3.8863

Model(s) Covered:

Escort 178

Source:

Escort 178 Instruction Manual

Compatibility:

MET/CAL 7.20 SP1 (7.20x) or later

Additional Equipment Required:

WaveTek 9105 Work Mat

The average run time during testing was 12 minutes (keyboard entry).