Package: escort_179_1yr_ver_5520.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Escort 179: (1 year) CAL VER /5520, Revision: 1.2
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Escort 179: (1 year) CAL VER /5520
Date: 2010-06-23 15:22:02
Revision: 1.2
Automated Reference: Fluke 5520A
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 5520A (1 year) specifications.
CAPACITANCE TESTS:
40 µF Range, 39.00µF; TUR 2.9764
Model(s) Covered:
Escort 179
Source:
Escort 179 Instruction Manual
Compatibility:
MET/CAL or 5500/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
Fluke 80AK SMP to double banana adapter.
TYPE K SMP plug to SMP plug thermocouple cables (1).
1. Included in 5500A/LEADS, Lead Set
The average run time during testing was 20 minutes (keyboard entry).