Package: escort_179_1yr_ver_5520.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Escort 179: (1 year) CAL VER /5520

Date: 2010-06-23 15:22:02

Revision: 1.2

Automated Reference: Fluke 5520A

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5520A (1 year) specifications.

CAPACITANCE TESTS:

40 µF Range, 39.00µF; TUR 2.9764

Model(s) Covered:

Escort 179

Source:

Escort 179 Instruction Manual

Compatibility:

MET/CAL or 5500/CAL 7.20 SP1 (7.20x) or later

Additional Equipment Required:

Fluke 80AK SMP to double banana adapter.

TYPE K SMP plug to SMP plug thermocouple cables (1).

1. Included in 5500A/LEADS, Lead Set

The average run time during testing was 20 minutes (keyboard entry).