Package: escort_3145a_1yr_ver_rs232_5520.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Escort 3145A: (1 year) CAL VER RS-232 /5520

Date: 2010-05-12 14:38:43

Revision: 1.1

Automated Reference: Fluke 5520A

Model(s) Covered:

Escort 3145A

Source:

ESCORT-3146A, ESCORT-3145A 5 1/2 - Digit Dual Display Multimeter

Operation Manual

Compatibility:

MET/CAL or 5500/CAL 7.20 SP3 (7.20Z) or later

Additional Equipment Required:

None

The average run time during testing was 20 minutes.