Escort 3146A: (1 year) CAL VER RS-232 /5520

Date: 2010-05-12 14:38:46

Revision: 1.1

Automated Reference: Fluke 5520A

Model(s) Covered:

Escort 3146A

Source:

ESCORT-3146A, ESCORT-3146A 5 1/2 - Digit Dual Display Multimeter

Operation Manual

Compatibility:

MET/CAL or 5500/CAL 7.20 SP3 (7.20Z) or later

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1 based on 5520A 1 year specifications.

DC CURRENT TESTS:

12 A Range

- 11.9 A; TUR 3.6875

Additional Equipment Required:

None

The average run time during testing was 20 minutes.