Package: extech_mt3x0_cal_5500.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

EXTECH Multitec 310: (1 year) CAL VER /5500

EXTECH Multitec 320: (1 year) CAL VER /5500

EXTECH Multitec 330: (1 year) CAL VER /5500

Date: 2003-06-03

Revision: 1.1

Automated Reference: Fluke 5500A

Model(s) Covered: MT310, MT320 MT330

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1 based on 5500A 90 day specifications:

RESISTANCE

- 39.00 MOhm; TUR 2.7245

FREQUENCY

- 4.900 Hz; TUR 3.2509

CAPACITANCE

- 39.00 µF; TUR 3.5455

Source: EXTECH Instruments

MT310, MT320, MT330

Specifications in User's Guide, Version 1.2 4/2002

Compatibility: MET/CAL or 5500/CAL 7.01 or later

Additional Equipment Required: None

The average run time during testing was 10 minutes.

EXTECH Multitec 310: CAL ADJ /5500

EXTECH Multitec 320: CAL ADJ /5500

EXTECH Multitec 330: CAL ADJ /5500

Date: 2003-06-03

Revision: 1.1

Automated Reference: Fluke 5500A

Model(s) Covered: MT310, MT320 MT330

Source: EXTECH Instruments

CALIBRATION PROCEDURE

MT310 Series of DMM's, V 1.0 3/26/02

MT320 Series of DMM's, V 1.0 3/26/02

MT330 Series of DMM's, V 1.0 3/26/02

Compatibility: MET/CAL or 5500/CAL 7.01 or later

Additional Equipment Required: None

The average run time during testing was 10 minutes.