Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Fluke 114: (1 year) CAL VER /5520, Revision: 1.1
Fluke 115/115C: (1 year) CAL VER /5520, Revision: 1.1
Fluke 116/116C: (1 year) CAL VER /5520, Revision: 1.1
Fluke 117/117C: (1 year) CAL VER /5520, Revision: 1.1
Fluke 114: CAL ADJ /5520, Revision: 1.4
Fluke 115/115C: CAL ADJ /5520, Revision: 1.4
Fluke 116/116C: CAL ADJ /5520, Revision: 1.4
Fluke 117/117C: CAL ADJ /5520, Revision: 1.4
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 114: (1 year) CAL VER /5520
Date: 2009-06-19 12:24:01
Revision: 1.1
Automated Reference: Fluke 5520A
Fluke 115/115C: (1 year) CAL VER /5520
Date: 2009-06-19 12:24:04
Revision: 1.1
Automated Reference: Fluke 5520A
Model(s) Covered:
115, 115C
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 5520A (1 year) specifications.
CAPACITANCE TESTS
9999 µF Range: 900 µF; TUR 3.7822
Compatibility:
MET/CAL or 5500/CAL SP1 (7.20x) or later
Additional Equipment Required:
None
The average run time during testing was 10 minutes.
Fluke 116/116C: (1 year) CAL VER /5520
Date: 2009-06-19 12:24:07
Revision: 1.1
Automated Reference: Fluke 5520A
Model(s) Covered:
116, 116C
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 5520A (1 year) specifications.
CAPACITANCE TESTS
9999 µF Range: 900 µF; TUR 3.7822
Compatibility:
MET/CAL or 5500/CAL SP1 (7.20x) or later
Additional Equipment Required:
Fluke 80AK K-Type DMM (thermocouple) Adapter
Type K thermocouple extension cable with SMP (m)-(m) plugs
The average run time during testing was 10 minutes.
Fluke 117/117C: (1 year) CAL VER /5520
Date: 2009-06-19 12:24:10
Revision: 1.1
Automated Reference: Fluke 5520A
Model(s) Covered:
117, 117C
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 5520A (1 year) specifications.
CAPACITANCE TESTS
9999 µF Range: 900 µF; TUR 3.7822
Compatibility:
MET/CAL or 5500/CAL SP1 (7.20x) or later
Additional Equipment Required:
Double Banana plug
The average run time during testing was 10 minutes.
Date: 2009-06-19 12:24:13
Revision: 1.4
Automated Reference: Fluke 5520A
Model(s) Covered:
114, 115, 115C, 116, 116C, 117, 117C
Compatibility:
MET/CAL or 5500/CAL SP1 (7.20x) or later
Additional Equipment Required:
None
The average run time during testing was 8 minutes.
Model(s) Covered:
114
Compatibility:
MET/CAL or 5500/CAL SP1 (7.20x) or later
Additional Equipment Required:
None
The average run time during testing was 10 minutes.