Fluke 12E: (1 year) CAL VER /9100 Chs

Date: 2010-04-02 07:48:08

Revision: 1.1

Automated Reference: Datron 9100

Model(s) Covered:

12E

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 9100 (1 year) specifications.

CAPACITANCE TEST

50nF Range:

48 nF, TUR 2.6316

500 nF Range:

480 nF, TUR 1.4205

5 µF Range:

4.8 µF TUR 3.0500

50 µF Range:

48 µF TUR 3.0500

DCI TEST

10 A Range

10.00 A, TUR 3.9130

AC mA TEST

400 mA Range

350 mA, TUR 3.3735

Compatibility:

MET/CAL or 5500/CAL 7.20 SP3 (7.20Z) or later

Additional Equipment Required:

None

The average run time during testing was: 10 minutes