Fluke 1503: (1 year) CAL VER /5320,5520,DMM, Revision: 2
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 1503: (1 year) CAL VER /5320,5520,DMM
Date: 2010-08-17 13:01:26
Revision: 2
Automated Reference: Digital Multimeter
Automated Reference: Fluke 5320A
Automated Reference: Fluke 5520A
Model(s) Covered:
1503
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 5320A (1 year) specifications.
SOURCE VOLTAGE ACCURACY TESTS, Open Circuit:
50 kOhm
50 V; TUR 3.9600
INSULATION
1900 MOhm, 1000 V; TUR 1.7409 (1503 only)
49 MOhm, 1000 V; TUR 3.1505
60 MOhm, 1000 V; TUR 2.9167
EARTH RESISTANCE ACCURACY TESTS
20 Ohm Range
2 Ohms; TUR 3.7500
Source:
Fluke 150x Insulation Testers, Calibration Manual
PN 2465477, November 2005, Rev. 1, 7/07
Compatibility:
MET/CAL 7.20p or later
Additional Equipment Required:
None
The average run time during testing was 15 minutes.