Package: fluke_1507_1yr_ver_5320_5520_dmm.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 1507: (1 year) CAL VER /5320,5520,DMM

Date: 2010-08-17 13:01:37

Revision: 2

Automated Reference: Digital Multimeter

Automated Reference: Fluke 5320A

Automated Reference: Fluke 5520A

Model(s) Covered:

1507

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5320A (1 year) specifications.

SOURCE VOLTAGE ACCURACY TESTS, Open Circuit:

50 kOhm

50 V; TUR 3.9600

INSULATION

1900 MOhm, 1000 V; TUR 1.7409 (1503 only)

49 MOhm, 1000 V; TUR 3.1505

60 MOhm, 1000 V; TUR 2.9167

EARTH RESISTANCE ACCURACY TESTS

20 Ohm Range

2 Ohms; TUR 3.7500

Source:

Fluke 150x Insulation Testers, Calibration Manual

PN 2465477, November 2005, Rev. 1, 7/07

Compatibility:

MET/CAL 7.20p or later

Additional Equipment Required:

None

The average run time during testing was 15 minutes.