Package: fluke_1555_1yr_ver_5320_vlc_5320load_pwr_res.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Fluke 1555: (1 year) CAL VER /5320+VLC,5320LOAD,PwrRes, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 1555: (1 year) CAL VER /5320+VLC,5320LOAD,PwrRes
Date: 2010-11-12 12:53:00
Revision: 1.1
Automated Reference: Fluke 5320A
Automated Reference: Fluke 80K-40
Manual Reference: 0.11 uF HV Test Fixture
Manual Reference: 1 uF HV Test Fixture
Manual Reference: 10MOhm ±1.25% 10kV resistor
Manual Reference: 1GOhm ±1.25% 10kV resistor
Manual Reference: 200GOhm ±1.25% 10kV resistor
Manual Reference: 2TOhm ±5% 10kV resistor
Manual Reference: Fluke 5320A-LOAD
Model(s) Covered:
1555
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 5320 (1 year) specifications.
INSULATION ACCURACY TESTS:
1000 V, 1 GOhm; TUR 3.85
2500 V, 1 GOhm; TUR 3.85
5000 V, 1 GOhm; TUR 3.85
OUTPUT VOLTAGE ACCURACY:
250 V, No Load; TUR 2.22
SHORT CIRCUIT CURRENT:
1.5 mA @ 5000 V; TUR 1.97
Source:
Fluke 1550C/1555 Insulation Tester
Calibration Manual
August 2010
Compatibility:
MET/CAL or 5500/CAL 7.3 or later
Additional Equipment Required:
IR Cable Assembly, Fluke P/N 1578406
10 kV Power Resistors
10 MOhm ±1.25%
1 GOhm ±1.25%
200 GOhm ±1.25%
2 TOhm ±5%
HV Capacitance Test Fixtures w/Bleeder Resistors:
8 ea. 8 µF capacitors in series.
Ducati 8 µF, 450 V motor start type AC 4.16.10.13.64
Farnell Stock No 1219055
3 ea. 0.33 µF capacitors in series.
Arcotronics MKP C4 0.33 µF 2000 V
RS Components stock No 384-7694
The average run time during testing was 20 minutes.