Fluke 1577: (1 year) CAL VER /5500,5320+VLC,80K-40,DMM

Date: 2009-12-29 07:22:43

Revision: 1.9

Automated Reference: Digital Multimeter

Automated Reference: Fluke 5320A

Automated Reference: Fluke 5500A

Automated Reference: Fluke 80K-40

Model(s) Covered:

1577

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5500A (1 year) specifications.

INSULATION RESISTANCE ACCURACY

1000 V Range

600 MOhm; TUR 2.9167

Source:

Fluke 1587/1577 Insulation Multimeters Calibration Manual

PN 2456783, September 2005

Fluke 1587/1577 Insulation Multimeters Users Manual

PN 2401027, April 2005, Rev. 1, 6/09

Compatibility:

MET/CAL or 5500/CAL 7.20 SP1 (7.20x) or later

Additional Equipment Required:

Probe, test, safety plug (normally supplied with UUT)

DC Power Supply 0~15 V

The average run time during testing was 20 minutes.

Fluke 1577: (1 year) CAL VER /5500,5320+VLC,DMM

Date: 2009-12-29 07:22:47

Revision: 1.3

Automated Reference: Digital Multimeter

Automated Reference: Fluke 5320A

Automated Reference: Fluke 5500A

Model(s) Covered:

1577

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5500A (1 year) specifications.

INSULATION RESISTANCE ACCURACY

1000 V Range

600 MOhm; TUR 2.9167

Source:

Fluke 1587/1577 Insulation Multimeters Calibration Manual

PN 2456783, September 2005

Fluke 1587/1577 Insulation Multimeters Users Manual

PN 2401027, April 2005, Rev. 1, 6/09

Compatibility:

MET/CAL or 5500/CAL 7.20 SP1 (7.20x) or later

Additional Equipment Required:

Probe, test, safety plug (normally supplied with UUT)

DC Power Supply 0~15 V

The average run time during testing was 20 minutes.

Fluke 1587: (1 year) CAL VER /5500,5320+VLC,80K-40,DMM

Date: 2009-12-29 07:22:50

Revision: 1.8

Automated Reference: Digital Multimeter

Automated Reference: Fluke 5320A

Automated Reference: Fluke 5500A

Automated Reference: Fluke 80K-40

Model(s) Covered:

1587

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5500A (1 year) specifications.

RESISTANCE

50 MOhm Range

50 MOhm; TUR 3.0528

CAPACITANCE TESTS

1000 nF Range

700 nF; TUR 3.7818

INSULATION RESISTANCE ACCURACY

1000 V Range

600 MOhm; TUR 2.9167

Source:

Fluke 1587/1577 Insulation Multimeters Calibration Manual

PN 2456783, September 2005

Fluke 1587/1577 Insulation Multimeters Users Manual

PN 2401027, April 2005, Rev .1, 6/09

Compatibility:

MET/CAL or 5500/CAL 7.20 SP1 (7.20x) or later

Additional Equipment Required:

Probe, test, safety plug (normally supplied with UUT)

Fluke 80AK K-Type (thermocouple) DMM Adapter

Type K thermocouple extension cable with SMP (m)-(m) plugs

DC Power Supply 0~15 V

The average run time during testing was 25 minutes.

Fluke 1587: (1 year) CAL VER /5500,5320+VLC,DMM

Date: 2009-12-29 07:22:53

Revision: 1.3

Automated Reference: Digital Multimeter

Automated Reference: Fluke 5320A

Automated Reference: Fluke 5500A

Model(s) Covered:

1587

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5500A (1 year) specifications.

RESISTANCE

50 MOhm Range

50 MOhm; TUR 3.0528

CAPACITANCE TESTS

1000 nF Range

700 nF; TUR 3.7818

INSULATION RESISTANCE ACCURACY

1000 V Range

600 MOhm; TUR 2.9167

Source:

Fluke 1587/1577 Insulation Multimeters Calibration Manual

PN 2456783, September 2005

Fluke 1587/1577 Insulation Multimeters Users Manual

PN 2401027, April 2005, Rev .1, 6/09

Compatibility:

MET/CAL or 5500/CAL 7.20 SP1 (7.20x) or later

Additional Equipment Required:

Probe, test, safety plug (normally supplied with UUT)

Fluke 80AK K-Type (thermocouple) DMM Adapter

Type K thermocouple extension cable with SMP (m)-(m) plugs

DC Power Supply 0~15 V

The average run time during testing was 25 minutes.

Fluke 1587T: (1 year) CAL VER /5500,5320+VLC,DMM

Date: 2009-12-29 07:22:56

Revision: 1.6

Automated Reference: Digital Multimeter

Automated Reference: Fluke 5320A

Automated Reference: Fluke 5500A

Model(s) Covered:

1587T

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5500A (1 year) specifications.

RESISTANCE

50 MOhm Range

50 MOhm; TUR 3.0528

CAPACITANCE TESTS

1000 nF Range

700 nF; TUR 3.7818

Source:

Fluke 1587/1577 Insulation Multimeters Calibration Manual

PN 2456783, September 2005

Fluke 1587/1577 Insulation Multimeters Users Manual

PN 2401027, April 2005, Rev .1, 6/09

Compatibility:

MET/CAL or 5500/CAL 7.20 SP1 (7.20x) or later

Additional Equipment Required:

Probe, test, safety plug (normally supplied with UUT)

Fluke 80AK K-Type (thermocouple) DMM Adapter

Type K thermocouple extension cable with SMP (m)-(m) plugs

DC Power Supply 0~15 V

The average run time during testing was 25 minutes.

Fluke 1577: CAL ADJ /5500

Fluke 1587: CAL ADJ /5500

Fluke 1587T: CAL ADJ /5500

Date: 2009-12-29 07:23:03

Revision: 1.3

Automated Reference: Fluke 5500A

Model(s) Covered:

1577, 1587, 1587T

Source:

Fluke 1587/1577 Insulation Multimeters Calibration Manual

PN 2456783, September 2005

Fluke 1587/1577 Insulation Multimeters Users Manual

PN 2401027, April 2005, Rev .1, 1/06

Compatibility:

MET/CAL or 5500/CAL 7.20 SP1 (7.20x) or later

Additional Equipment Required:

None

The average run time during testing was 10 minutes.