Fluke 1577: (1 year) CAL VER /5520,5320+VLC,80K-40,DMM, Revision: 1.7
Fluke 1577: (1 year) CAL VER /5520,5320+VLC,DMM, Revision: 1.3
Fluke 1587: (1 year) CAL VER /5520,5320+VLC,80K-40,DMM, Revision: 1.7
Fluke 1587: (1 year) CAL VER /5520,5320+VLC,DMM, Revision: 1.3
Fluke 1587T: (1 year) CAL VER /5520,5320+VLC,DMM, Revision: 1.6
Fluke 1577: CAL ADJ /5520, Revision: 1.2
Fluke 1587: CAL ADJ /5520, Revision: 1.2
Fluke 1587T: CAL ADJ /5520, Revision: 1.2
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 1577: (1 year) CAL VER /5520,5320+VLC,80K-40,DMM
Date: 2009-12-29 07:33:23
Revision: 1.7
Automated Reference: Digital Multimeter
Automated Reference: Fluke 5320A
Automated Reference: Fluke 5520A
Automated Reference: Fluke 80K-40
Fluke 1577: (1 year) CAL VER /5520,5320+VLC,DMM
Date: 2009-12-29 07:33:27
Revision: 1.3
Automated Reference: Digital Multimeter
Automated Reference: Fluke 5320A
Automated Reference: Fluke 5520A
Model(s) Covered:
1577
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 5520A (1 year) specifications.
INSULATION RESISTANCE ACCURACY
1000 V Range
600 MOhm; TUR 2.9167
Source:
Fluke 1587/1577 Insulation Multimeters Calibration Manual
PN 2456783, September 2005
Fluke 1587/1577 Insulation Multimeters Users Manual
PN 2401027, April 2005, Rev. 1, 6/09
Compatibility:
MET/CAL or 5520/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
Probe, test, safety plug (normally supplied with UUT)
DC Power Supply 0~15 V
The average run time during testing was 20 minutes.
Fluke 1587: (1 year) CAL VER /5520,5320+VLC,80K-40,DMM
Date: 2009-12-29 07:33:30
Revision: 1.7
Automated Reference: Digital Multimeter
Automated Reference: Fluke 5320A
Automated Reference: Fluke 5520A
Automated Reference: Fluke 80K-40
Model(s) Covered:
1587
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 5520A (1 year) specifications.
RESISTANCE
50 MOhm Range
50 MOhm; TUR 3.0528
CAPACITANCE TESTS
1000 nF Range
700 nF; TUR 3.7818
INSULATION RESISTANCE ACCURACY
1000 V Range
600 MOhm; TUR 2.9167
Source:
Fluke 1587/1577 Insulation Multimeters Calibration Manual
PN 2456783, September 2005
Fluke 1587/1577 Insulation Multimeters Users Manual
PN 2401027, April 2005, Rev .1, 6/09
Compatibility:
MET/CAL or 5520/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
Probe, test, safety plug (normally supplied with UUT)
Fluke 80AK K-Type (thermocouple) DMM Adapter
Type K thermocouple extension cable with SMP (m)-(m) plugs
DC Power Supply 0~15 V
The average run time during testing was 25 minutes.
Fluke 1587: (1 year) CAL VER /5520,5320+VLC,DMM
Date: 2009-12-29 07:33:33
Revision: 1.3
Automated Reference: Digital Multimeter
Automated Reference: Fluke 5320A
Automated Reference: Fluke 5520A
Model(s) Covered:
1587
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 5520A (1 year) specifications.
RESISTANCE
50 MOhm Range
50 MOhm; TUR 3.0528
CAPACITANCE TESTS
1000 nF Range
700 nF; TUR 3.7818
INSULATION RESISTANCE ACCURACY
1000 V Range
600 MOhm; TUR 2.9167
Source:
Fluke 1587/1577 Insulation Multimeters Calibration Manual
PN 2456783, September 2005
Fluke 1587/1577 Insulation Multimeters Users Manual
PN 2401027, April 2005, Rev .1, 6/09
Compatibility:
MET/CAL or 5520/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
Probe, test, safety plug (normally supplied with UUT)
Fluke 80AK K-Type (thermocouple) DMM Adapter
Type K thermocouple extension cable with SMP (m)-(m) plugs
DC Power Supply 0~15 V
The average run time during testing was 25 minutes.
Fluke 1587T: (1 year) CAL VER /5520,5320+VLC,DMM
Date: 2009-12-29 07:33:36
Revision: 1.6
Automated Reference: Digital Multimeter
Automated Reference: Fluke 5320A
Automated Reference: Fluke 5520A
Model(s) Covered:
1587T
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 5520A (1 year) specifications.
RESISTANCE
50 MOhm Range
50 MOhm; TUR 3.0528
CAPACITANCE TESTS
1000 nF Range
700 nF; TUR 3.7818
Source:
Fluke 1587/1577 Insulation Multimeters Calibration Manual
PN 2456783, September 2005
Fluke 1587/1577 Insulation Multimeters Users Manual
PN 2401027, April 2005, Rev .1, 6/09
Compatibility:
MET/CAL or 5520/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
Probe, test, safety plug (normally supplied with UUT)
Fluke 80AK K-Type (thermocouple) DMM Adapter
Type K thermocouple extension cable with SMP (m)-(m) plugs
DC Power Supply 0~15 V
The average run time during testing was 25 minutes.
Date: 2009-12-29 07:33:44
Revision: 1.2
Automated Reference: Fluke 5520A
Model(s) Covered:
1577, 1587, 1587T
Source:
Fluke 1587/1577 Insulation Multimeters Calibration Manual
PN 2456783, September 2005
Fluke 1587/1577 Insulation Multimeters Users Manual
PN 2401027, April 2005, Rev .1, 1/06
Compatibility:
MET/CAL or 5520/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
None
The average run time during testing was 10 minutes.
Model(s) Covered:
1577
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 5520A (1 year) specifications.
INSULATION RESISTANCE ACCURACY
1000 V Range
600 MOhm; TUR 2.9167
Source:
Fluke 1587/1577 Insulation Multimeters Calibration Manual
PN 2456783, September 2005
Fluke 1587/1577 Insulation Multimeters Users Manual
PN 2401027, April 2005, Rev. 1, 6/09
Compatibility:
MET/CAL or 5520/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
Probe, test, safety plug (normally supplied with UUT)
DC Power Supply 0~15 V
The average run time during testing was 20 minutes.