Package: fluke_15b_1yr_ver_9100.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 15B: (1 year) CAL VER /9100

Date: 2010-03-11 13:46:34

Revision: 1.1

Automated Reference: Datron 9100

Model(s) Covered:

15B

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 9100 (1 year) specifications.

CAPACITANCE TEST

50nF Range:

48 nF, TUR 3.3224

500 nF Range:

480 nF, TUR 2.8693

Compatibility:

MET/CAL 7.20 SP3 (7.20Z) or later

The average run time during testing was: 10 minutes