Fluke 1577: (1 year) CAL VER /9100,189,80K-6

Fluke 1587: (1 year) CAL VER /9100,189,80K-6

Date: 2006-04-04 07:51:52

Revision: 1.3

Automated Reference: Datron 9100

Manual Reference: Fluke 189

Manual Reference: Fluke 80K-6

Model(s) Covered:

Fluke 1577, 1587

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 9100 (1 year) specifications.

DC MILLIAMPERE TESTS

600 mA Range

400 mA DC; TUR 2.7933

CAPACITANCE TESTS

1000 nF Range

700 nF; TUR 2.2727

9999 µF Range

8000 µF; TUR 1.3286

INSULATION TESTS

500 V Range

450 Mohm; TUR 3.8095

1000 V Range

600 Mohm; TUR 3.3333

Source:

Fluke 1587/1577 Insulation Multimeters Calibration

PN 2456783, September 2005

Compatibility:

MET/CAL 7.11 SP1 (7.11j) or later

Additional Equipment Required:

Probe, test, safety plug (normally supplied with UUT)

Fluke 80AK K-Type (thermocouple) DMM Adapter

Type K thermocouple extension cable with SMP (m)-(m) plugs

DC Power Supply 0~15 V

The average run time during testing was 25 minutes (keyboard entry).

Fluke 1577: CAL ADJ /9100

Fluke 1587: CAL ADJ /9100

Date: 2006-04-04 07:51:56

Revision: 1.1

Automated Reference: Datron 9100

Model(s) Covered:

Fluke 1577, 1587

Source:

Fluke 1587/1577 Insulation Multimeters Calibration

PN 2456783, September 2005

Compatibility:

MET/CAL 7.11 SP1 (7.11j) or later

Additional Equipment Required:

None

The average run time during testing was 10 minutes.