Fluke 1652B: (1 year) CAL VER RS-232 /5320+VLC,ResBox

Date: 2012-02-15 06:27:16

Revision: 2.7

Automated Reference: Fluke 5320A

Manual Reference: Decade "R" Box

Model Covered:

1652B w/Firmware: UI >= V3.14, Analog >= V2.33

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1.

Insulation

- 18 MOhm, 250 V; TUR 3.3333

- 180 MOhm, 250 V; TUR 3.3333

- 18 MOhm, 500 V; TUR 3.3333

- 180 MOhm, 500 V; TUR 3.3333

- 180 MOhm, 1000 V; TUR 2.0833

Continuity

- 2 Ohm; TUR 3.7500

Loop

- 50 Ohm; Comp; TUR 3.3333

- 180 Ohm; Comp; TUR 3.6000

Source:

Fluke 1651B, 1652B, 1653B

Electrical Installation Tester

Product Specifications

Preliminary Calibration Information, July 2008

Fluke 1652C/1653B/1654B

Electrical Installation Tester

Calibration Manual

March 2011

Manual Supplement

Manual Title: 1652C/1653B/1654B Calibration Supplement Issue: 2

Part Number: Web-Only

Issue Date: 1/12

Source Author: Fluke

Compatibility: 5500/CAL or MET/CAL 7.20C or later

Additional Equipment Required:

Decade Resistor Box

Resistance: 18, 180, & 1800 Ohms

Accuracy: 0.75 % or better

Power: 0.5 W minimum

Fluke IR Adapter-Serial Cable, Part No. 2043365

Fluke-165xB, Zero Adapter, Part No. 3301338