Package: fluke_1653_1yr_ver_rs232_5320_vlc.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 1653: (1 year) CAL VER RS-232 /5320+VLC

Date: 2011-04-20 09:14:42

Revision: 2

Automated Reference: Fluke 5320A

Model Covered: 1653

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1.

Insulation

- 10 MOhm, 250 V; TUR 3.6000

- 180 MOhm, 250 V; TUR 3.3333

- 10 MOhm, 500 V; TUR 3.6000

- 180 MOhm, 500 V; TUR 3.3333

- 180 MOhm, 1000 V; TUR 2.0833

Continuity

- 2 Ohm; TUR 1.9355

Loop

- 1.8 Ohm; Comp; TUR 3.0196

- 18 Ohm; Comp; TUR 2.1695

- 50 Ohm; Comp; TUR 1.8405

- 180 Ohm; Comp; TUR 1.9183

- 1 kOhm; Comp; TUR 3.9960

RCD (Trip Current)

- 10 mA x5 Sinusoid; TUR 2.0000

Source: Fluke 165x Factory Test Procedure

Source Author: Fluke

Compatibility: 5500/CAL or MET/CAL 7.20p or later

Additional Equipment Required:

IR Adapter-Serial Cable, Part No. 2043365