Package: fluke_1653_1yr_ver_rs232_5320_vlc.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Fluke 1653: (1 year) CAL VER RS-232 /5320+VLC, Revision: 2
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 1653: (1 year) CAL VER RS-232 /5320+VLC
Date: 2011-04-20 09:14:42
Revision: 2
Automated Reference: Fluke 5320A
Model Covered: 1653
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1.
Insulation
- 10 MOhm, 250 V; TUR 3.6000
- 180 MOhm, 250 V; TUR 3.3333
- 10 MOhm, 500 V; TUR 3.6000
- 180 MOhm, 500 V; TUR 3.3333
- 180 MOhm, 1000 V; TUR 2.0833
Continuity
- 2 Ohm; TUR 1.9355
Loop
- 1.8 Ohm; Comp; TUR 3.0196
- 18 Ohm; Comp; TUR 2.1695
- 50 Ohm; Comp; TUR 1.8405
- 180 Ohm; Comp; TUR 1.9183
- 1 kOhm; Comp; TUR 3.9960
RCD (Trip Current)
- 10 mA x5 Sinusoid; TUR 2.0000
Source: Fluke 165x Factory Test Procedure
Source Author: Fluke
Compatibility: 5500/CAL or MET/CAL 7.20p or later
Additional Equipment Required:
IR Adapter-Serial Cable, Part No. 2043365