Package: fluke_16_1yr_ver_5500.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 16: (1 year) CAL VER /5500

Date: 2012-08-07 13:02:04

Revision: 1.4

Automated Reference: Fluke 5500A

Model(s) Covered:

Fluke 16

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5500A (1 year) specifications.

AC CURRENT TESTS

200 µA Range: 40.0 µA @ 50 Hz; TUR 3.8462

200 µA Range: 40.0 µA @ 400 Hz; TUR 3.8462

RESISTANCE TESTS

40 MOhm Range: 35.0 MOhm; TUR 3.8144

Compatibility:

MET/CAL or 5500/CAL SP1 (7.20x) or later

Additional Equipment Required:

Fluke 80AK K-Type DMM (thermocouple) Adapter

Type K thermocouple extension cable with SMP (m)-(m) plugs

The average run time during testing was 13 minutes.