Package: fluke_17_19_cal_5700_5725_5520.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 17: (1 yr) CAL VER /5700,5725,5520

Date: 2009-10-19 10:32:57

Revision: 1.1

Automated Reference: Fluke 5520A

Automated Reference: Fluke 5700A

Automated Reference: Fluke 5725A

Model(s) Covered:

17

This procedure contains the following test(s) whose Test Uncertainty Ratio

are less than 4:1.

Capacitance Test:

5 nF:

4.70 nF; TUR 3.5894

Source:

Fluke Model 17 & 19 Multimeters Calibration Information

PN 690826 November 1998

Compatibility:

MET/CAL 7.20 SP1 (7.20x) or later

Additional Equipment Required:

None

The average run time during testing was 6 minutes or less.

Fluke 19: (1 yr) CAL VER /5700,5725,5520

Date: 2009-10-19 10:33:01

Revision: 1.1

Automated Reference: Fluke 5520A

Automated Reference: Fluke 5700A

Automated Reference: Fluke 5725A

Model(s) Covered:

19

This procedure contains the following test(s) whose Test Uncertainty Ratio

are less than 4:1.

Capacitance Test:

5 nF:

4.70 nF; TUR 3.5894

Source:

Fluke Model 17 & 19 Multimeters Calibration Information

PN 690826 November 1998

Compatibility:

MET/CAL 7.20 SP1 (7.20x) or later

Additional Equipment Required:

None

The average run time during testing was 6 minutes or less.