Package: fluke_183_185_rs232_9100.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 183: (1 year) CAL VER RS-232 /9100

Fluke 185: (1 year) CAL VER RS-232 /9100

Date: 2004-12-10

Revision: 1.2

Automated Reference: Datron 9100

Model(s) Covered:

Fluke 183, 185

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 9100 (1 year) specifications.

AC VOLTAGE TEST

- 4.6 V, 20 kHz; TUR 3.5377 (183)

- 4.6 V, 20 kHz; TUR 2.3585 (185)

FREQUENCY TEST

- 990 kHz, 2.828Vp; TUR 1.2121

CAPACITANCE TEST

- 4 nF; TUR 1.6667

- 40 nF; TUR 2.8667

- 400 nF; TUR 3.1618

- 4 µF; TUR 2.4432

- 40 mF; TUR 2.6739

DC CURRENT TEST

- 460 mA; TUR 2.8426

AC CURRENT TEST

- 34 µA, 1 kHz; TUR 1.2353

- 460 mA, 60 Hz; TUR 3.1489

- 4.6 A, 60 Hz; TUR 2.4262

- 10 A, 60 Hz; TUR 3.4783

Source:

Fluke 183/185 Calibration Manual, Part No. 1610436, May 2001,

Rev. 1, 6/02, Manual Supplement 1, Issue Date 3/04

Compatibility:

MET/CAL 7.01 SP2 or later

Additional Equipment Required:

DB9 IR Cable Kit, Fluke Part No. 1590638

Wavetek 9105 Work Mat

Fluke 183: CAL ADJ RS-232 /9100

Fluke 185: CAL ADJ RS-232 /9100

Date: 2004-12-10

Revision: 1.2

Automated Reference: Datron 9100

Models covered:

Fluke 183, Fluke 185

Source:

Fluke 183/185 Calibration Manual, Part No. 1610436, May 2001,

Rev. 1, 6/02, Manual Supplement 1, Issue Date 3/04

Compatibility:

MET/CAL 7.01 SP2 or later

Required Files:

metfir.exe (custom executable)

Additional Equipment Required:

DB9 IR Cable Kit, Fluke Part No. 1590638

Wavetek 9105 Work Mat